Streamlined pin-ridge-filter design for single-energy proton FLASH planning.
Med Phys
; 51(4): 2955-2966, 2024 Apr.
Article
em En
| MEDLINE
| ID: mdl-38214381
ABSTRACT
BACKGROUND:
FLASH radiotherapy (FLASH-RT) with ultra-high dose rate has yielded promising results in reducing normal tissue toxicity while maintaining tumor control. Planning with single-energy proton beams modulated by ridge filters (RFs) has been demonstrated feasible for FLASH-RT.PURPOSE:
This study explored the feasibility of a streamlined pin-shaped RF (pin-RF) design, characterized by coarse resolution and sparsely distributed ridge pins, for single-energy proton FLASH planning.METHODS:
An inverse planning framework integrated within a treatment planning system was established to design streamlined pin RFs for single-energy FLASH planning. The framework involves generating a multi-energy proton beam plan using intensity-modulated proton therapy (IMPT) planning based on downstream energy modulation strategy (IMPT-DS), followed by a nested pencil-beam-direction-based (PBD-based) spot reduction process to iteratively reduce the total number of PBDs and energy layers along each PBD for the IMPT-DS plan. The IMPT-DS plan is then translated into the pin-RFs and the single-energy beam configurations for IMPT planning with pin-RFs (IMPT-RF). This framework was validated on three lung cases, quantifying the FLASH dose of the IMPT-RF plan using the FLASH effectiveness model. The FLASH dose was then compared to the reference dose of a conventional IMPT plan to measure the clinical benefit of the FLASH planning technique.RESULTS:
The IMPT-RF plans closely matched the corresponding IMPT-DS plans in high dose conformity (conformity index of <1.2), with minimal changes in V7Gy and V7.4 Gy for the lung (<3%) and small increases in maximum doses (Dmax) for other normal structures (<3.4 Gy). Comparing the FLASH doses to the doses of corresponding IMPT-RF plans, drastic reductions of up to nearly 33% were observed in Dmax for the normal structures situated in the high-to-moderate-dose regions, while negligible changes were found in Dmax for normal structures in low-dose regions. Positive clinical benefits were seen in comparing the FLASH doses to the reference doses, with notable reductions of 21.4%-33.0% in Dmax for healthy tissues in the high-dose regions. However, in the moderate-to-low-dose regions, only marginal positive or even negative clinical benefit for normal tissues were observed, such as increased lung V7Gy and V7.4 Gy (up to 17.6%).CONCLUSIONS:
A streamlined pin-RF design was developed and its effectiveness for single-energy proton FLASH planning was validated, revealing positive clinical benefits for the normal tissues in the high dose regions. The coarsened design of the pin-RF demonstrates potential advantages, including cost efficiency and ease of adjustability, making it a promising option for efficient production.Palavras-chave
Texto completo:
1
Base de dados:
MEDLINE
Assunto principal:
Radioterapia de Intensidade Modulada
/
Terapia com Prótons
/
Neoplasias
Limite:
Humans
Idioma:
En
Revista:
Med Phys
Ano de publicação:
2024
Tipo de documento:
Article
País de afiliação:
Estados Unidos