Heart rate sensitivity of virtual non-contrast calcium scores derived from photon counting detector CT data: a phantom study.
Risch, Franka; Schwarz, Florian; Kroencke, Thomas; Decker, Josua A.
Afiliação
  • Risch F; Department of Diagnostic and Interventional Radiology, University Hospital Augsburg, Stenglinstr. 2, 86156, Augsburg, Germany.
  • Schwarz F; Department of Diagnostic and Interventional Radiology, University Hospital Augsburg, Stenglinstr. 2, 86156, Augsburg, Germany.
  • Kroencke T; Medical Faculty, Ludwig Maximilian University Munich, Munich, Germany.
  • Decker JA; Clinic for Diagnostic and Interventional Radiology, Donau-Isar-Klinikum, Deggendorf, Germany.
Radiol Med ; 129(3): 401-410, 2024 Mar.
Article em En | MEDLINE | ID: mdl-38319495
ABSTRACT

PURPOSE:

To assess the reliability of virtual non-contrast (VNC) derived coronary artery calcium quantities in relation to heart rate and the VNC algorithm used compared to reference true non-contrast (TNC), considering several clinically established acquisition modes. MATERIAL AND

METHODS:

An ad hoc built coronary phantom containing four calcified lesions and an iodinated lumen was scanned using three cardiac acquisition modes three times within an anthropomorphic cardiac motion phantom simulating different heart rates (0, 60, 80, 100 bpm) and reconstructed with a conventional (VNCconv) and a calcium-sensitive (VNCpc) VNC algorithm. TNC reference was scanned at 0 bpm with non-iodinated lumen. Calcium scores were assessed in terms of number of lesions detected, Agatston and volume scores and global noise was measured. Paired t-test and Wilcoxon test were performed to test measurements for significant difference.

RESULTS:

For both VNC algorithms used, calcium levels or noise were not significantly affected by heart rate. Measurements on VNCpc reconstructions best reproduced TNC results, but with increased variability (Agatston scores at 0 bpm for TNC, VNCconv, and VNCpc were 47.1 ± 1.1, 6.7 ± 2.8 (p < 0.001), and 45.3 ± 7.6 (p > 0.05), respectively). VNC reconstructions showed lower noise levels compared to TNC, especially for VNCpc (noiseheart on TNC, VNCconv and VNCpc at 0 bpm was 5.0 ± 0.4, 4.5 ± 0.2, 4.2 ± 0.2).

CONCLUSION:

No significant heart rate dependence of VNC-based calcium scores was observed in an intra-reconstruction comparison. VNCpc reproduces TNC scores better than VNCconv without significant differences and decreased noise, however, with an increasing average deviation with rising heart rates. VNC-based CACS should be used with caution as the measures show higher variability compared to reference TNC and therefore hold the potential of incorrect risk categorization.