Optimal oversizing in transcatheter aortic valve replacement with the self-expanding Evolut valve system.
Mas-Peiro, Silvia; Alperi, Alberto; Avvedimento, Marisa; Regueiro, Ander; Avanzas, Pablo; Angellotti, Domenico; Esposito, Giovanni; Vidal-Cales, Pablo; Mohammadi, Siamak; Farjat-Pasos, Julio; Moris, Cesar; Rodés-Cabau, Josep.
Afiliação
  • Mas-Peiro S; Quebec Heart and Lung Institute, Laval University, Quebec City, Quebec, Canada.
  • Alperi A; Cardiology Department, Hospital Universitario Central de Asturias, Oviedo, Spain.
  • Avvedimento M; Quebec Heart and Lung Institute, Laval University, Quebec City, Quebec, Canada; Department of Advanced Biomedical Sciences, Federico II University of Naples, Naples, Italy.
  • Regueiro A; Cardiology Department, Instituto Clínic Cardiovascular, Hospital Clínic, Barcelona, Spain and Institut d'Investigacions Biomèdiques August Pi i Sunyer (IDIBAPS), Barcelona, Spain.
  • Avanzas P; Cardiology Department, Hospital Universitario Central de Asturias, Oviedo, Spain.
  • Angellotti D; Department of Advanced Biomedical Sciences, Federico II University of Naples, Naples, Italy.
  • Esposito G; Department of Advanced Biomedical Sciences, Federico II University of Naples, Naples, Italy.
  • Vidal-Cales P; Cardiology Department, Instituto Clínic Cardiovascular, Hospital Clínic, Barcelona, Spain and Institut d'Investigacions Biomèdiques August Pi i Sunyer (IDIBAPS), Barcelona, Spain.
  • Mohammadi S; Quebec Heart and Lung Institute, Laval University, Quebec City, Quebec, Canada.
  • Farjat-Pasos J; Quebec Heart and Lung Institute, Laval University, Quebec City, Quebec, Canada.
  • Moris C; Cardiology Department, Hospital Universitario Central de Asturias, Oviedo, Spain.
  • Rodés-Cabau J; Quebec Heart and Lung Institute, Laval University, Quebec City, Quebec, Canada; Cardiology Department, Instituto Clínic Cardiovascular, Hospital Clínic, Barcelona, Spain and Institut d'Investigacions Biomèdiques August Pi i Sunyer (IDIBAPS), Barcelona, Spain. E-mail: josep.rodes@criucpq.ulaval.ca.
J Invasive Cardiol ; 36(6)2024 Jun.
Article em En | MEDLINE | ID: mdl-38446025
ABSTRACT

OBJECTIVES:

Valve oversizing has been associated with reduced paravalvular leaks (PVL) and valve migration risk. However, no optimal cut-off oversizing value has been defined for the Evolut system (Medtronic). The aim of this study was to assess the relationship between the degree of oversizing and moderate-to-severe PVL and determine the optimal oversizing cut-off value.

METHODS:

We conducted a multicenter study that included 740 consecutive patients with multidetector computed tomography (CT) data. Valve size was selected according to manufacturer recommendations, with oversizing ranging from 10% to 30%. The primary endpoint was moderate-to-severe PVL.

RESULTS:

The median age was 84 years (79-87 years), with 58.4% women, and a median EuroSCORE II of 4.1% (2.4-7.3%). Moderate-to-severe PVL was observed in 7.0% of the patients. An inverse relationship was found between oversizing and both PVL (11.3%, 8.6%, 5.4%, and 2.7% for quartiles Q1 to Q4; P = .007) and the need for post-dilation (P = .016). The multivariable analysis showed an association between oversizing and PVL (OR 0.915 for each 1%-increase, P = .002). The optimal oversizing cut-off value to predict PVL was 20%, and PVL was significantly higher in patients with oversizing less than 20% (10.5% vs.4.2%, P less than .001). There were no differences in major clinical events according to the degree of oversizing, and a higher oversizing did not translate into an increased risk of permanent pacemaker (18.4% vs18.3%, P = .976).

CONCLUSIONS:

In TAVR with the Evolut valve, a higher oversizing was associated with lower rates of moderate-to-severe PVL and a lower need for post-dilation, with no negative impact on procedural and early clinical outcomes. A 20% oversizing threshold could be suggested to reduce PVLs. Further prospective studies are warranted to validate optimal oversizing for this valve system.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Valva Aórtica / Estenose da Valva Aórtica / Desenho de Prótese / Próteses Valvulares Cardíacas / Tomografia Computadorizada Multidetectores / Substituição da Valva Aórtica Transcateter Limite: Aged / Aged80 / Female / Humans / Male Idioma: En Revista: J Invasive Cardiol Assunto da revista: CARDIOLOGIA Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Canadá

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Valva Aórtica / Estenose da Valva Aórtica / Desenho de Prótese / Próteses Valvulares Cardíacas / Tomografia Computadorizada Multidetectores / Substituição da Valva Aórtica Transcateter Limite: Aged / Aged80 / Female / Humans / Male Idioma: En Revista: J Invasive Cardiol Assunto da revista: CARDIOLOGIA Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Canadá