X-ray beam diagnostics at the MID instrument of the European X-ray Free-Electron Laser Facility.
J Synchrotron Radiat
; 31(Pt 3): 596-604, 2024 May 01.
Article
em En
| MEDLINE
| ID: mdl-38587894
ABSTRACT
The Materials Imaging and Dynamics (MID) instrument at the European X-ray Free-Electron Laser Facility (EuXFEL) is equipped with a multipurpose diagnostic end-station (DES) at the end of the instrument. The imager unit in DES is a key tool for aligning the beam to a standard trajectory and for adjusting optical elements such as focusing lenses or the split-and-delay line. Furthermore, the DES features a bent-diamond-crystal spectrometer to disperse the spectrum of the direct beam to a line detector. This enables pulse-resolved characterization of the EuXFEL spectrum to provide X-ray energy calibration, and the spectrometer is particularly useful in commissioning special modes of the accelerator. Together with diamond-based intensity monitors, the imager and spectrometer form the DES unit which also contains a heavy-duty beamstop at the end of the MID instrument. Here, we describe the setup in detail and provide exemplary beam diagnostic results.
Texto completo:
1
Base de dados:
MEDLINE
Idioma:
En
Revista:
J Synchrotron Radiat
Assunto da revista:
RADIOLOGIA
Ano de publicação:
2024
Tipo de documento:
Article
País de afiliação:
Alemanha