Anatomic fitting studies of a total artificial heart in heart transplant recipients. Critical dimensions and prediction of fit.
ASAIO J
; 42(5): M337-42, 1996.
Article
em En
| MEDLINE
| ID: mdl-8944902
Anatomic fitting studies of the Cleveland Clinic-Nimbus total artificial heart were performed in 33 patients undergoing heart transplantation. The pump fit in the pericardial space in 20 men (80%) and 4 women (50%). There was no significant difference between the Fit and Non-Fit groups in external chest dimensions. Among 42 intrathoracic dimensions, the distance from the center of the mitral valve to the diaphragm (Fit: 5.6 +/- 2.2 cm, Non-Fit: 3.6 +/- 0.4 cm, p < 0.00001) and the distance from the caudal end of the pulmonary valve to the diaphragm (Fit: 9.4 +/- 1.6 cm, Non-Fit: 6.3 +/- 0.8 cm, p < 0.0001) were the most critical. To predict anatomic fit, an index (A x B x C) was obtained from chest X-ray measurements (A, the craniocaudal distance from the dorsal region of the 8th left rib to the left diaphragm; B, the maximum left chest width; and C, the maximum anteroposterior sternum-vertebrae dimension). The pump fit in 88.5% of the patients with an index above 1200 cm3, whereas it fit in only 14.3% of the patients with an index below 1200 cm3 (p < 0.001). This index was an easily obtainable, good predictor of anatomic fit.
Buscar no Google
Base de dados:
MEDLINE
Assunto principal:
Transplante de Coração
/
Coração Artificial
Tipo de estudo:
Diagnostic_studies
/
Evaluation_studies
/
Prognostic_studies
/
Risk_factors_studies
Limite:
Female
/
Humans
/
Male
Idioma:
En
Revista:
ASAIO J
Assunto da revista:
TRANSPLANTE
Ano de publicação:
1996
Tipo de documento:
Article
País de afiliação:
Estados Unidos