Resultados 1 - 16 de 16
Filtrar
1.
Circ Arrhythm Electrophysiol ; 17(1): e012026, 2024 01.
Artigo em Inglês | MEDLINE | ID: mdl-38152949

RESUMO

BACKGROUND: Effects of contact force (CF) on lesion formation during pulsed field ablation (PFA) have not been well validated. The purpose of this study was to determine the relationship between average CF and lesion size during PFA using a swine-beating heart model. METHODS: A 7F catheter with a 3.5-mm ablation electrode and CF sensor (TactiCath SE, Abbott) was connected to a PFA system (CENTAURI, Galvanize Therapeutics). In 5 closed-chest swine, biphasic PFA current was delivered between the ablation electrode and a skin patch at 40 separate sites in right ventricle (28 Amp) and 55 separate sites in left ventricle (35 Amp) with 4 different levels of CF: (1) low (CF range of 4-13 g; median, 9.5 g); (2) moderate (15-30 g; median, 21.5 g); (3) high (34-55 g; median, 40 g); and (4) no electrode contact, 2 mm away from the endocardium. Swine were sacrificed at 2 hours after ablation, and lesion size was measured using triphenyl tetrazolium chloride staining. In 1 additional swine, COX (cytochrome c oxidase) staining was performed to examine mitochondrial activity to delineate reversible and irreversible lesion boundaries. Histological examination was performed with hematoxylin and eosin and Masson trichrome staining. RESULTS: Ablation lesions were well demarcated with triphenyl tetrazolium chloride staining, showing (1) a dark central zone (contraction band necrosis and hemorrhage); (2) a pale zone (no mitochondrial activity and nuclear pyknosis, indicating apoptosis zone); and a hyperstained zone by triphenyl tetrazolium chloride and COX staining (unaffected normal myocardium with preserved mitochondrial activity, consistent with reversible zone). At constant PFA current intensity, lesion depth increased significantly with increasing CF. There were no detectable lesions resulting from ablation without electrode contact. CONCLUSIONS: Acute PFA ventricular lesions show irreversible and reversible lesion boundaries by triphenyl tetrazolium chloride staining. Electrode-tissue contact is required for effective lesion formation during PFA. At the same PFA dose, lesion depth increases significantly with increasing CF.


Assuntos
Ablação por Cateter , Ventrículos do Coração , Suínos , Animais , Ventrículos do Coração/cirurgia , Ventrículos do Coração/patologia , Ablação por Cateter/efeitos adversos , Ablação por Cateter/métodos , Cloretos , Coração , Catéteres
2.
JACC Clin Electrophysiol ; 9(9): 1854-1863, 2023 09.
Artigo em Inglês | MEDLINE | ID: mdl-37480857

RESUMO

BACKGROUND: Pulsed electrical field (PEF) ablation may cause tissue heating. These changes are reportedly small, but each PEF system and waveform will have a different behavior, and data are lacking. OBJECTIVES: This study sought to compare the temperature profile of focal point, monopolar biphasic PEF ablation versus radiofrequency (RF). METHODS: Ablation lesions were performed on perfused thigh muscle of swine. PEF lesions were performed with 3 compatible ablation catheters at the highest (25 amp) energy, and 1 catheter (Tacticath SE) was also used at the 22- and 19-amp levels. Temperature changes in the tissue were measured using fluoroptic temperature probes inserted at the muscle surface, as well as 3 mm and 7 mm below the surface. Temperatures were recorded continuously at baseline, during delivery, and after ablation. Muscle temperatures were compared with those of RF lesions performed with 1 catheter (Tacticath SE) at 30 W for 30 seconds. RESULTS: PEF ablation with 3energy settings produced small temperature changes. Maximum average temperature rise for PEF for the maximum (25-amp) energy setting (32 lesions) was 7.6 °C, 2.8 °C, and 0.9 °C at the surface, 3-mm depth, and 7-mm depth, respectively. The temperature rise was dose dependent, with lower energy settings yielding less temperature rise. RF ablations (10 lesions) produced temperature increases of 16.6 °C, 39.8 °C, and 9.5 °C at the surface, 3-mm depth, and 7-mm depth, respectively. CONCLUSIONS: PEF caused detectable temperature changes in muscle tissue, which never exceeded 2.8 °C at the 3-mm depth versus baseline. By contrast, RF produced substantial temperature rises. These data support that focal monopolar biphasic energy delivered by this PEF technology retains a favorable thermal safety profile.


Assuntos
Ablação por Cateter , Animais , Suínos , Catéteres , Eletricidade , Terapia de Eletroporação Irreversível , Temperatura
3.
Artigo em Inglês | MEDLINE | ID: mdl-37188871

RESUMO

BACKGROUND: Pulsed field ablation (PFA) may have a superior safety profile compared to other technologies, but it has the potential to cause gaseous microbubbles (MB), which may be associated with cerebral emboli. Limited relative safety data has been published regarding PFA in the left ventricle (LV). METHODS: Healthy and chronic myocardial infarction (MI) swine underwent PFA (monopolar, biphasic, 25 Amps) in the LV using an irrigated focal catheter under intra-cardiac echocardiography (ICE) guidance for MB monitoring. Two control swine received air MBs through the lumen of the ablation catheter. Swine underwent brain MRI before and after PFA (or control air MB injection). Gross pathology and histology of brains with abnormal MRI findings were performed. RESULTS: Four healthy and 5 chronic MI swine underwent 124 left ventricular PFA applications. No PFA-related MB formation was noted on ICE. Both control swine developed multiple acute emboli in the thalamus and caudate on DWI, ADC, and FLAIR brain MRI images in response to air MB injection. Of the 9 PFA swine, there were no abnormalities on ADC or FLAIR images. There was one hyperintense focus in the left putamen on the DWI trace image, but the absence of ADC or FLAIR affirmation suggested it was artifact. Gross pathology and histopathology of this region did not detect any abnormalities. CONCLUSIONS: Focal monopolar biphasic PFA of both healthy and chronically infarcted left ventricular myocardium does not generate any MB or cerebral emboli observable on ICE and brain MRI.

4.
Europace ; 25(4): 1503-1509, 2023 04 15.
Artigo em Inglês | MEDLINE | ID: mdl-36793229

RESUMO

AIMS: Data on ventricular pulsed-field ablation (PFA) are sparse in the setting of chronic myocardial infarction (MI). The objective of this study was to compare the biophysical and histopathologic characteristics of PFA in healthy and MI swine ventricular myocardium. METHODS AND RESULTS: Myocardial infarction swine (n = 8) underwent coronary balloon occlusion and survived for 30 days. We then performed endocardial unipolar, biphasic PFA of the MI border zone and a dense scar with electroanatomic mapping and using an irrigated contact force (CF)-sensing catheter with the CENTAURI System (Galaxy Medical). Lesion and biophysical characteristics were compared with three controls: MI swine undergoing thermal ablation, MI swine undergoing no ablation, and healthy swine undergoing similar PFA applications that included linear lesion sets. Tissues were systematically assessed by gross pathology utilizing 2,3,5-triphenyl-2H-tetrazolium chloride staining and histologically with haematoxylin and eosin and trichrome. Pulsed-field ablation in healthy myocardium generated well-demarcated ellipsoid lesions (7.2 ± 2.1 mm depth) with contraction band necrosis and myocytolysis. Pulsed-field ablation in MI demonstrated slightly smaller lesions (depth 5.3 ± 1.9 mm, P = 0.0002), and lesions infiltrated into the irregular scar border, resulting in contraction band necrosis and myocytolysis of surviving myocytes and extending to the epicardial border of the scar. Coagulative necrosis was present in 75% of thermal ablation controls but only in 16% of PFA lesions. Linear PFA resulted in contiguous linear lesions with no gaps in gross pathology. Neither CF nor local R-wave amplitude reduction correlated with lesion size. CONCLUSION: Pulsed-field ablation of a heterogeneous chronic MI scar effectively ablates surviving myocytes within and beyond the scar, demonstrating promise for the clinical ablation of scar-mediated ventricular arrhythmias.


Assuntos
Ablação por Cateter , Infarto do Miocárdio , Taquicardia Ventricular , Suínos , Animais , Cicatriz , Miocárdio/patologia , Ventrículos do Coração , Coração , Arritmias Cardíacas , Ablação por Cateter/métodos
5.
Cell Rep ; 42(2): 112123, 2023 02 28.
Artigo em Inglês | MEDLINE | ID: mdl-36795562

RESUMO

The contribution of cross-presenting XCR1+ dendritic cells (DCs) and SIRPα+ DCs in maintaining T cell function during exhaustion and immunotherapeutic interventions of chronic infections remains poorly characterized. Using the mouse model of chronic LCMV infection, we found that XCR1+ DCs are more resistant to infection and highly activated compared with SIRPα+ DCs. Exploiting XCR1+ DCs via Flt3L-mediated expansion or XCR1-targeted vaccination notably reinvigorates CD8+ T cells and improves virus control. Upon PD-L1 blockade, XCR1+ DCs are not required for the proliferative burst of progenitor exhausted CD8+ T (TPEX) cells but are indispensable to sustain the functionality of exhausted CD8+ T (TEX) cells. Combining anti-PD-L1 therapy with increased frequency of XCR1+ DCs improves functionality of TPEX and TEX subsets, while increase of SIRPα+ DCs dampened their proliferation. Together, this demonstrates that XCR1+ DCs are crucial for the success of checkpoint inhibitor-based therapies through differential activation of exhausted CD8+ T cell subsets.


Assuntos
Apresentação Cruzada , Viroses , Camundongos , Animais , Células Dendríticas , Linfócitos T CD8-Positivos , Imunoterapia
6.
J Cardiovasc Electrophysiol ; 34(1): 99-107, 2023 01.
Artigo em Inglês | MEDLINE | ID: mdl-36335638

RESUMO

OBJECTIVES: Pulsed electric field (PEF) therapies employ punctuated energy delivery to kill cells in a volume of tissue through mechanisms that are not dependent on thermal processes. A key component to successful cardiac ablation procedures is ensuring the generation of transmural, contiguous ablation zones, which requires in-depth knowledge regarding treatment sizes for a given therapeutic application. METHODS: In this study, a series of acute treatments were delivered to porcine ventricles, where triphenyl tetrazolium chloride (TTC) vitality stain was used to identify treatment effect sizes for the three focal monopolar CENTAURI PEF cardiac ablation energy settings. RESULTS: Treatment depths were 5.7, 7.2, and 8.2 mm for the 19, 22, and 25 A energy settings, respectively. Gross pathology indicated umbral zones of hemorrhage surrounded by pale avital TTC-negative-negative tissue, which contrasted significantly from radiofrequency ablation (RF) controls. Histologically, treatment zones are identified by regions of contraction band necrosis and cardiomyocytolysis, which contrasted with RF control lesions composed primarily of coagulation necrosis. CONCLUSIONS: Together, these data indicate the ability for focal monopolar PEF treatments to generate deep treatment zones in cardiac ablation without incurring the gross or histological coagulative characteristics of RF thermal lesions.


Assuntos
Ablação por Cateter , Ablação por Radiofrequência , Suínos , Animais , Ablação por Cateter/efeitos adversos , Ablação por Cateter/métodos , Necrose , Catéteres , Ventrículos do Coração
7.
Card Electrophysiol Clin ; 14(4): 757-767, 2022 12.
Artigo em Inglês | MEDLINE | ID: mdl-36396191

RESUMO

Current ablation systems rely on thermal energy to produce ablation lesions (heating: RF, laser and ultrasound, and cooling: cryo-thermia). While thermal ablation has been proven to be effective, there are several limitations: 1) relatively long procedural times; 2) high recurrence rate of ventricular arrhythmias; and 3) excessive heating potentially leading to serious complications, including steam pop (perforation), coronary arterial injury and thrombo-embolism. Pulsed field ablation (PFA)/irreversible electroporation (IRE) offers a unique non-thermal ablation strategy which has the potential to overcome these limitations. Recent pre-clinical studies suggest that PFA/IRE might be effective and safe for the treatment of cardiac arrhythmias.


Assuntos
Ablação por Cateter , Criocirurgia , Humanos , Arritmias Cardíacas , Coração , Vasos Coronários
8.
Circ Arrhythm Electrophysiol ; 14(9): e010086, 2021 09.
Artigo em Inglês | MEDLINE | ID: mdl-34538095

RESUMO

Pulsed electrical field (PEF) energy is a promising technique for catheter ablation of cardiac arrhythmias. In this article, the key aspects that need to be considered for safe and effective PEF delivery are reviewed, and their impact on clinical feasibility is discussed. The most important benefit of PEF appears to be the ability to kill cells through mechanisms that do not alter stromal proteins, sparing sensitive structures to improve safety, without sacrificing cardiomyocyte ablation efficacy. Many parameters affect PEF treatment outcomes, including pulse intensity, waveform shape, and number of pulses, as well as electrode configuration and geometry. These physical and electrical characteristics must be titrated carefully to balance target tissue effects with collateral implications (muscle contraction, temperature rise, risk of electrical arcing events). It is important to note that any combination of parameters affecting PEF needs to be tested for clinical efficacy and safety. Applying PEF clinically requires knowledge of the fundamentals of this technology to exploit its opportunities and generate viable, durable health improvements for patients.


Assuntos
Fibrilação Atrial/cirurgia , Ablação por Cateter/métodos , Eletroporação/métodos , Veias Pulmonares/cirurgia , Humanos , Resultado do Tratamento
9.
IEEE Trans Biomed Circuits Syst ; 14(4): 867-878, 2020 08.
Artigo em Inglês | MEDLINE | ID: mdl-32746346

RESUMO

Sensing implants that can be deployed by catheterization or by injection are preferable over implants requiring invasive surgery. However, present powering methods for active implants and present interrogation methods for passive implants require bulky parts within the implants that hinder the development of such minimally invasive devices. In this article, we propose a novel approach that potentially enables the development of passive sensing systems overcoming the limitations of previous implantable sensing systems in terms of miniaturization. In this approach implants are shaped as thread-like devices suitable for implantation by injection. Their basic structure consists of a thin elongated body with two electrodes at opposite ends and a simple and small circuit made up of a diode, a capacitor and a resistor. The interrogation method to obtain measurements from the implants consists in applying innocuous bursts of high frequency (≥1 MHz) alternating current that reach the implants by volume conduction and in capturing and processing the voltage signals that the implants produce after the bursts. As proof-of-concept, and for illustrating how to put in practice this novel approach, here we describe the development and characterization of a system for measuring the conductivity of tissues surrounding the implant. We also describe the implementation and the in vitro validation of a 0.95 mm-thick, flexible injectable implant made of off-the-shelf components. For conductivities ranging from about 0.2 to 0.8 S/m, when compared to a commercial conductivity meter, the accuracy of the implemented system was about ±10%.


Assuntos
Condutividade Elétrica , Miniaturização/instrumentação , Monitorização Fisiológica/instrumentação , Próteses e Implantes , Eletrônica Médica , Humanos , Perna (Membro)/fisiologia , Músculo Esquelético/fisiologia , Desenho de Prótese , Transdutores
10.
Ann Biomed Eng ; 48(5): 1451-1462, 2020 May.
Artigo em Inglês | MEDLINE | ID: mdl-32026232

RESUMO

High-frequency irreversible electroporation (H-FIRE) has emerged as an alternative to conventional irreversible electroporation (IRE) to overcome the issues associated with neuromuscular electrical stimulation that appear in IRE treatments. In H-FIRE, the monopolar pulses typically used in IRE are replaced with bursts of short bipolar pulses. Currently, very little is known regarding how the use of a different waveform affects the cell death dynamics and mechanisms. In this study, human pancreatic adenocarcinoma cells were treated with a typical IRE protocol and various H-FIRE schemes with the same energized time. Cell viability, membrane integrity and Caspase 3/7 activity were assessed at different times after the treatment. In both treatments, we identified two different death dynamics (immediate and delayed) and we quantified the electric field ranges that lead to each of them. While in the typical IRE protocol, the electric field range leading to a delayed cell death is very narrow, this range is wider in H-FIRE and can be increased by reducing the pulse length. Membrane integrity in cells suffering a delayed cell death shows a similar time evolution in all treatments, however, Caspase 3/7 expression was only observed in cells treated with H-FIRE.


Assuntos
Morte Celular , Eletroporação/métodos , Caspase 3/metabolismo , Caspase 7/metabolismo , Técnicas de Cultura de Células , Linhagem Celular Tumoral , Humanos
11.
Artigo em Inglês | MEDLINE | ID: mdl-28557354

RESUMO

Metastases in the liver frequently grow as scattered tumor nodules that neither can be removed by surgical resection nor focally ablated. Previously, we have proposed a novel technique based on irreversible electroporation that may be able to simultaneously treat all nodules in the liver while sparing healthy tissue. The proposed technique requires increasing the electrical conductivity of healthy liver by injecting a hypersaline solution through the portal vein. Aiming to assess the capability of increasing the global conductivity of the liver by means of hypersaline fluids, here, it is presented a mathematical model that estimates the NaCl distribution within the liver and the resulting conductivity change. The model fuses well-established compartmental pharmacokinetic models of the organ with saline injection models used for resuscitation treatments, and it considers changes in sinusoidal blood viscosity because of the hypertonicity of the solution. Here, it is also described a pilot experimental study in pigs in which different volumes of NaCl 20% (from 100 to 200 mL) were injected through the portal vein at different flow rates (from 53 to 171 mL/minute). The in vivo conductivity results fit those obtained by the model, both quantitatively and qualitatively, being able to predict the maximum conductivity with a 14.6% average relative error. The maximum conductivity value was 0.44 second/m, which corresponds to increasing 4 times the mean basal conductivity (0.11 second/m). The results suggest that the presented model is well suited for predicting on liver conductivity changes during hypertonic saline injection.


Assuntos
Fígado/metabolismo , Animais , Condutividade Elétrica , Fígado/química , Modelos Teóricos , Solução Salina/química
12.
Phys Med Biol ; 63(3): 035027, 2018 01 31.
Artigo em Inglês | MEDLINE | ID: mdl-29235992

RESUMO

Electroporation-based treatments typically consist of the application of high-voltage dc pulses. As an undesired side effect, these dc pulses cause electrical stimulation of excitable tissues such as motor nerves. The present in vivo study explores the use of bursts of sinusoidal voltage in a frequency range from 50 kHz to 2 MHz, to induce irreversible electroporation (IRE) whilst avoiding neuromuscular stimulation. A series of 100 dc pulses or sinusoidal bursts, both with an individual duration of 100 µs, were delivered to rabbit liver through thin needles in a monopolar electrode configuration, and thoracic movements were recorded with an accelerometer. Tissue samples were harvested three hours after treatment and later post-processed to determine the dimensions of the IRE lesions. Thermal damage due to Joule heating was ruled out via computer simulations. Sinusoidal bursts with a frequency equal to or above 100 kHz did not cause thoracic movements and induced lesions equivalent to those obtained with conventional dc pulses when the applied voltage amplitude was sufficiently high. IRE efficacy dropped with increasing frequency. For 100 kHz bursts, it was estimated that the electric field threshold for IRE is about 1.4 kV cm-1 whereas that of dc pulses is about 0.5 kV cm-1.


Assuntos
Estimulação Elétrica/efeitos adversos , Eletroporação/métodos , Fígado/efeitos da radiação , Músculos/efeitos da radiação , Fibras Nervosas/efeitos da radiação , Órgãos em Risco/efeitos da radiação , Animais , Fígado/patologia , Modelos Teóricos , Músculos/patologia , Fibras Nervosas/patologia , Coelhos
13.
Radiol Oncol ; 51(4): 415-421, 2017 Dec.
Artigo em Inglês | MEDLINE | ID: mdl-29333120

RESUMO

BACKGROUND: Spread hepatic tumours are not suitable for treatment either by surgery or conventional ablation methods. The aim of this study was to evaluate feasibility and safety of selectively increasing the healthy hepatic conductivity by the hypersaline infusion (HI) through the portal vein. We hypothesize this will allow simultaneous safe treatment of all nodules by irreversible electroporation (IRE) when applied in a transhepatic fashion. MATERIAL AND METHODS: Sprague Dawley (Group A, n = 10) and Athymic rats with implanted hepatic tumour (Group B, n = 8) were employed. HI was performed (NaCl 20%, 3.8 mL/Kg) by trans-splenic puncture. Deionized serum (40 mL/Kg) and furosemide (2 mL/Kg) were simultaneously infused through the jugular vein to compensate hypernatremia. Changes in conductivity were monitored in the hepatic and tumour tissue. The period in which hepatic conductivity was higher than tumour conductivity was defined as the therapeutic window (TW). Animals were monitored during 1-month follow-up. The animals were sacrificed and selective samples were used for histological analysis. RESULTS: The overall survival rate was 82.4% after the HI protocol. The mean maximum hepatic conductivity after HI was 2.7 and 3.5 times higher than the baseline value, in group A and B, respectively. The mean maximum hepatic conductivity after HI was 1.4 times higher than tumour tissue in group B creating a TW to implement selective IRE. CONCLUSIONS: HI through the portal vein is safe when the hypersaline overload is compensated with deionized serum and it may provide a TW for focused IRE treatment on tumour nodules.

14.
J Neural Eng ; 12(6): 066010, 2015 Dec.
Artigo em Inglês | MEDLINE | ID: mdl-26447945

RESUMO

OBJECTIVE: It is possible to develop implantable microstimulators whose actuation principle is based on rectification of high-frequency (HF) current bursts supplied through skin electrodes. This has been demonstrated previously by means of devices consisting of a single diode. However, previous single diode devices caused dc currents which made them impractical for clinical applications. Here flexible thread-like stimulation implants which perform charge balance are demonstrated in vivo. APPROACH: The implants weigh 40.5 mg and they consist of a 3 cm long tubular silicone body with a diameter of 1 mm, two electrodes at opposite ends, and, within the central section of the body, an electronic circuit made up of a diode, two capacitors, and a resistor. In the present study, each implant was percutaneously introduced through a 14 G catheter into either the gastrocnemius muscle or the cranial tibial muscle of a rabbit hindlimb. Then stimulation was performed by delivering HF bursts (amplitude <60 V, frequency 1 MHz, burst repetition frequency from 10 Hz to 200 Hz, duration = 200 µs) through a pair of textile electrodes strapped around the hindlimb and either isometric plantarflexion or dorsiflexion forces were recorded. Stimulation was also assayed 1, 2 and 4 weeks after implantation. MAIN RESULTS: The implants produced bursts of rectified current whose mean value was of a few mA and were capable of causing local neuromuscular stimulation. The implants were well-tolerated during the 4 weeks. SIGNIFICANCE: Existing power supply methods, and, in particular inductive links, comprise stiff and bulky parts. This hinders the development of minimally invasive implantable devices for neuroprostheses based on electrical stimulation. The proposed methodology is intended to relieving such bottleneck. In terms of mass, thinness, and flexibility, the demonstrated implants appear to be unprecedented among the intramuscular stimulation implants ever assayed in vertebrates.


Assuntos
Eletrodos Implantados , Epiderme/fisiologia , Animais , Estimulação Elétrica/instrumentação , Estimulação Elétrica/métodos , Membro Posterior/fisiologia , Masculino , Microeletrodos , Músculo Esquelético/fisiologia , Coelhos
15.
Bioelectrochemistry ; 105: 16-24, 2015 Oct.
Artigo em Inglês | MEDLINE | ID: mdl-25955102

RESUMO

Delivery of the so-called Tumor Treatment Fields (TTFields) has been proposed as a cancer therapy. These are low magnitude alternating electric fields at frequencies from 100 to 300 kHz which are applied continuously in a non-invasive manner. Electric field delivery may produce an increase in temperature which cannot be neglected. We hypothesized that the reported results obtained by applying TTFields in vivo could be due to heat rather than to electrical forces as previously suggested. Here, an in vivo study is presented in which pancreatic tumors subcutaneously implanted in nude mice were treated for a week either with mild hyperthermia (41 °C) or with TTFields (6 V/cm, 150 kHz) and tumor growth was assessed. Although the TTFields applied singly did not produce any significant effect, the combination with chemotherapy did show a delay in tumor growth in comparison to animals treated only with chemotherapy (median relative reduction=47%). We conclude that concomitant chemotherapy and TTFields delivery show a beneficial impact on pancreatic tumor growth. Contrary to our hypothesis, this impact is non-related with the induced temperature increase.


Assuntos
Eletricidade , Neoplasias/patologia , Animais , Divisão Celular , Camundongos , Camundongos Nus
16.
Int J Hyperthermia ; 29(3): 211-8, 2013 May.
Artigo em Inglês | MEDLINE | ID: mdl-23573935

RESUMO

PURPOSE: The aim of this study was to assess the feasibility of a hybrid ablative technique based on applying electroporation (EP) pulses just before conducting radiofrequency ablation (RFA). The rationale was that the EP-induced reduction in blood perfusion could be sufficient to reduce the thermal sink effect and hence to increase the coagulation volume in comparison to that created exclusively by RFA. MATERIALS AND METHODS: A modelling study and in vivo experimental study were used. A Cool-tip RF applicator was used both for EP and RFA. RESULTS: Overall, the results did not show any synergy effect from using the hybrid technique. Applying EP pulses prior to RFA did not increase the coagulation zone obtained and the lesions were almost identical. Additional computer simulations provided an explanation for this; the effect of reducing blood perfusion by thermal damage during RFA completely masks the effect of reducing blood perfusion by EP. This is because both thermal damage and EP affect the same zone, i.e. the tissue around the electrode. CONCLUSIONS: Our computer modelling and in vivo experimental findings suggest that the combination of EP and RFA with monopolar applicators does not provide an additional benefit over the use of RFA alone.


Assuntos
Ablação por Cateter , Eletroporação , Modelos Teóricos , Animais , Terapia Combinada , Simulação por Computador , Estudos de Viabilidade , Feminino , Fígado/cirurgia , Suínos