AJR Am J Roentgenol ; 222(6): e2430988, 2024 Jun.
Artigo em Inglês | MEDLINE | ID: mdl-38506540

RESUMO

BACKGROUND. The energy demand of interventional imaging systems has historically been estimated using manufacturer-provided specifications rather than directly measured. OBJECTIVE. The purpose of this study was to investigate the energy consumption of interventional imaging systems and estimate potential savings in the carbon emissions and electricity costs of such systems through hypothetical operational adjustments. METHODS. An interventional radiology suite, neurointerventional suite, radiology fluoroscopy unit, two cardiology laboratories, and two urology fluoroscopy units were equipped with power sensors. Power measurement logs were extracted for a single 4-week period for each radiology and cardiology system (all between June 1, 2022, and November 28, 2022) and for the 2-week period from July 31, 2023, to August 13, 2023, for each urology system. Power statuses, procedure time stamps, and fluoroscopy times were extracted from various sources. System activity was divided into off, idle (no patient in room), active (patient in room for procedure), and net-imaging (active fluoroscopic image acquisition) states. Projected annual energy consumption was calculated. Potential annual savings in carbon emissions and electricity costs through hypothetical operational adjustments were estimated using published values for Switzerland. RESULTS. Across the seven systems, the mean power draw was 0.3-1.1, 0.7-7.4, 0.9-7.6, and 1.9-12.5 kW in the off, idle, active, and net-imaging states, respectively. Across systems, the off state, in comparison with the idle state, showed a decrease in the mean power draw of 0.2-6.9 kW (relative decrease, 22.2-93.2%). The systems had a combined projected annual energy consumption of 115,684 kWh (range, 3646-26,576 kWh per system). The systems' combined projected energy consumption occurring outside the net-imaging state accounted for 93.3% (107,978/115,684 kWh) of projected total energy consumption (range, 89.2-99.4% per system). A hypothetical operational adjustment whereby all systems would be switched from the idle state to the off state overnight and on weekends (versus being operated in idle mode 24 hours a day, 7 days a week) would yield the following potential annual savings: for energy consumption, 144,640 kWh; for carbon emissions, 18.6 metric tons of CO2 equivalent; and for electricity costs, US$37,896. CONCLUSION. Interventional imaging systems are energy intensive, having high consumption outside of image acquisition periods. CLINICAL IMPACT. Strategic operational adjustments (e.g., powering down idle systems) can substantially decrease the carbon emissions and electricity costs of interventional imaging systems.


Assuntos
Radiografia Intervencionista , Humanos , Radiografia Intervencionista/economia , Fluoroscopia/economia , Urologia/economia , Cardiologia/economia , Eletricidade , Pegada de Carbono
2.
Radiology ; 295(3): 593-605, 2020 06.
Artigo em Inglês | MEDLINE | ID: mdl-32208096

RESUMO

Background Awareness of energy efficiency has been rising in the industrial and residential sectors but only recently in the health care sector. Purpose To measure the energy consumption of modern CT and MRI scanners in a university hospital radiology department and to estimate energy- and cost-saving potential during clinical operation. Materials and Methods Three CT scanners, four MRI scanners, and cooling systems were equipped with kilowatt-hour energy measurement sensors (2-Hz sampling rate). Energy measurements, the scanners' log files, and the radiology information system from the entire year 2015 were analyzed and segmented into scan modes, as follows: net scan (actual imaging), active (room time), idle, and system-on and system-off states (no standby mode was available). Per-examination and peak energy consumption were calculated. Results The aggregated energy consumption imaging 40 276 patients amounted to 614 825 kWh, dedicated cooling systems to 492 624 kWh, representing 44.5% of the combined consumption of 1 107 450 kWh (at a cost of U.S. $199 341). This is equivalent to the usage in a town of 852 people and constituted 4.0% of the total yearly energy consumption at the authors' hospital. Mean consumption per CT examination over 1 year was 1.2 kWh, with a mean energy cost (±standard deviation) of $0.22 ± 0.13. The total energy consumption of one CT scanner for 1 year was 26 226 kWh ($4721 in energy cost). The net consumption per CT examination over 1 year was 3580 kWh, which is comparable to the usage of a two-person household in Switzerland; however, idle state consumption was fourfold that of net consumption (14 289 kWh). Mean MRI consumption over 1 year was 19.9 kWh per examination, with a mean energy cost of $3.57 ± 0.96. The mean consumption for a year in the system-on state was 82 174 kWh per MRI examination and 134 037 kWh for total consumption, for an energy cost of $24 127. Conclusion CT and MRI energy consumption is substantial. Considerable energy- and cost-saving potential is present during nonproductive idle and system-off modes, and this realization could decrease total cost of ownership while increasing energy efficiency. © RSNA, 2020.


Assuntos
Conservação de Recursos Energéticos/economia , Redução de Custos/economia , Imageamento por Ressonância Magnética/economia , Radiologia/economia , Tomografia Computadorizada por Raios X/economia , Alemanha , Humanos , Sistemas de Informação em Radiologia , Suíça