Biomed Tech (Berl) ; 64(5): 495-506, 2019 Sep 25.
Artigo em Inglês | MEDLINE | ID: mdl-30917101

RESUMO

In this review, scientific investigations of point-of-care testing (POCT) and point-of-care monitoring (POCM) devices are summarized with regard to the measurement accuracy of the hemoglobin concentration. As a common basis, information according to the Bland and Altman principle [bias, limits of agreement (LOA)] as well as the measurement accuracy and precision are considered, so that the comparability can be mapped. These collected data are subdivided according to the manufacturers, devices and procedures (invasive and non-invasive). A total of 31 devices were identified. A comparability of the scientific investigations in particular was given for 23 devices (18 invasive and five non-invasive measuring devices). In terms of measurement accuracy, there is a clear leap between invasive and non-invasive procedures, while no discernible improvement can be derived in the considered time frame from 2010 to 2018. According to the intended use, strict specifications result from the clinical standards, which are insufficiently met by the systems. More stringent requirements can be derived both in the area of blood donation and in the treatment of patients.


Assuntos
Hemoglobinas/química , Testes Imediatos , Hemoglobinas/análise , Humanos , Sistemas Automatizados de Assistência Junto ao Leito