Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros

Base de dados
Ano de publicação
Tipo de documento
Intervalo de ano de publicação
1.
J Soc Cardiovasc Angiogr Interv ; 2(6Part A): 101134, 2023.
Artigo em Inglês | MEDLINE | ID: mdl-39129905

RESUMO

Background: We sought to investigate the impact of stenting on native patent ductus arteriosus (PDA) length, curvature, and pulsatile deformations in patients with ductal-dependent pulmonary circulations. Methods: Patients with PDA stents who received contrast-enhanced 3-dimensional computed tomography with a view of the PDA, thoracic aorta, and pulmonary arteries were retrospectively included in this study. Geometric models of the prestented and poststented PDA were constructed from the computed tomography images, and PDA arclength, curvature, and pulsatile deformations were quantified. Results: A total of 12 patients with cyanotic congenital heart disease were included, 10 of whom received 1 stent in the PDA and 2 received multiple overlapping stents. From prestenting to poststenting, the PDA shortened by 26 ± 18% (P = .004) and decreased in mean and peak curvature by 60 ± 21% and 68 ± 15%, respectively (both P < .001). Pulsatile deformations varied highly for the native PDA, stented PDA, and stents themselves. Conclusions: The shortening and straightening of the PDA after stenting are significant and substantial, and their quantitative characterization will enable interventionalists to select stent lengths that span the entire PDA without encroaching on the aortic or pulmonary artery, which could cause hemodynamic interference, stent kink, and fatigue. Pulsatile PDA deformations can be used to design and evaluate devices tailored to congenital heart disease in neonates.

SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA