Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros

Base de dados
Ano de publicação
Tipo de documento
Intervalo de ano de publicação
1.
J Chromatogr A ; 1727: 464989, 2024 Jul 19.
Artigo em Inglês | MEDLINE | ID: mdl-38763085

RESUMO

Ultrahigh-performance liquid chromatography coupled with high-field quadrupole Orbitrap high resolution mass spectrometry was used for the separation and determination of 20 antihistamines, and a dispersive micro solid-phase extraction procedure using high-performance absorbing material was developed as a sample preparation strategy for extracting 20 antihistamines from milk. Instrument conditions and key parameters influencing extraction efficiency were investigated to obtain an optimized method. The limit of detection for 20 antihistamines in milk using this method is 0.05 µg/L to 1.0 µg/L. Recoveries are between 80.7 % and 108.3 %, and the relative standard deviation is less than 15 %. It is suitable for confirmatory monitoring and quantitative analysis of 20 antihistamines in milk. The results show that antihistamines in milk may be noteworthy issues for human health and environmental pollution.


Assuntos
Antagonistas dos Receptores Histamínicos , Limite de Detecção , Leite , Cromatografia Líquida de Alta Pressão/métodos , Leite/química , Animais , Antagonistas dos Receptores Histamínicos/análise , Antagonistas dos Receptores Histamínicos/isolamento & purificação , Microextração em Fase Sólida/métodos , Espectrometria de Massas/métodos , Bovinos , Reprodutibilidade dos Testes
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA