Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros

Base de dados
Ano de publicação
Tipo de documento
Intervalo de ano de publicação
1.
Clin Lab ; 65(6)2019 Jun 01.
Artigo em Inglês | MEDLINE | ID: mdl-31232031

RESUMO

BACKGROUND: We investigated a high-throughput and high-precision forward ABO blood typing screening method that utilizes a general-purpose biochemical analyzer to perform direct red blood cell sampling. METHODS: The blood group antisera used were Ortho® BioClone® Anti-A Serum and Ortho® BioClone® Anti-B Se-rum. AFFIRMAGEN® Reagent Red Blood Cells (Ortho Clinical Diagnostics) were used for AB standard red blood cells. The general-purpose biochemical analyzer employed was the TBATM-120FR HbA1c measurement unit (Canon Medical Systems). RESULTS: ABO blood group of patient samples was determined based on values relative to amount of change in the AFFIRMAGEN® response. Repeatability was CV5% or lower, and testing of 1,112 patient samples showed 100% agreement between the results obtained using the proposed method and those obtained using the tube test method. CONCLUSIONS: The proposed method allows ABO blood typing to be performed simply, quickly, and with a high degree of precision.


Assuntos
Sistema ABO de Grupos Sanguíneos/sangue , Automação Laboratorial , Tipagem e Reações Cruzadas Sanguíneas/instrumentação , Tipagem e Reações Cruzadas Sanguíneas/métodos , Eritrócitos , Hemoglobinas Glicadas/análise , Humanos , Reprodutibilidade dos Testes
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA