RESUMO
OBJECTIVE: The Woven EndoBridge (WEB) device was approved to treat wide-necked bifurcation aneurysms. The device is designed as an intrasaccular flow disruptor covering aneurysm widths up to 10 mm. Although prior studies combined all aneurysm sizes, it is known that aneurysms behave differently in response to endovascular treatment based on their size. Therefore, the authors' objective was to identify ideal middle cerebral artery (MCA) aneurysm width and neck sizes most suitable for WEB treatment. METHODS: The WorldWideWEB consortium is a large multicenter retrospective database that analyzes intracranial aneurysms treated with the WEB device. In this study, all unruptured MCA bifurcation aneurysms with available measurements were included. Cutoff values based on aneurysm width and neck in relation to aneurysm occlusion status were measured using the receiver operating characteristic (ROC) curve. Propensity score matching (PSM) was then used to compare treatment outcomes between aneurysms smaller and larger than the cutoff value for both width and neck size. RESULTS: The ideal cutoff values for MCA bifurcation aneurysm width and neck were 6.1 mm and 4.6 mm, respectively. On PSM, 87 matched pairs were compared based on width size (≤ 6.1 mm and > 6.1 mm), and 77 matched pairs were compared based on neck size (≤ 4.6 mm and > 4.6 mm). There was a significant difference in adequate aneurysm occlusion between aneurysms smaller and larger than those cutoff values for both widths (93% vs 76%, p = 0.0017) and neck sizes (90% vs 70%, p = 0.0026). The retreatment rate was also significantly higher for larger aneurysms in both parameters. CONCLUSIONS: This study shows that MCA bifurcation aneurysms ≤ 6.1 mm in width and ≤ 4.6 mm in neck size are significantly better candidates for WEB treatment, leading to improved occlusion status and reduced retreatment rate, which are important considerations when using WEB devices.
RESUMO
BACKGROUND: Endovascular treatment with the woven endobridge (WEB) device has been widely utilized for managing intracranial aneurysms. However, predicting the probability of achieving adequate occlusion (Raymond-Roy classification 1 or 2) remains challenging. OBJECTIVE: Our study sought to develop and validate a predictive calculator for adequate occlusion using the WEB device via data from a large multi-institutional retrospective cohort. METHODS: We used data from the WorldWide WEB Consortium, encompassing 356 patients from 30 centers across North America, South America, and Europe. Bivariate and multivariate regression analyses were performed on a variety of demographic and clinical factors, from which predictive factors were selected. Calibration and validation were conducted, with variance inflation factor (VIF) parameters checked for collinearity. RESULTS: A total of 356 patients were included: 124 (34.8%) were male, 108 (30.3%) were elderly (≥65 years), and 118 (33.1%) were current smokers. Mean maximum aneurysm diameter was 7.09 mm (SD 2.71), with 112 (31.5%) having a daughter sac. In the multivariate regression, increasing aneurysm neck size (OR 0.706 [95% CI: 0.535-0.929], p = 0.13) and partial aneurysm thrombosis (OR 0.135 [95% CI: 0.024-0.681], p = 0.016) were found to be the only statistically significant variables associated with poorer likelihood of achieving occlusion. The predictive calculator shows a c-statistic of 0.744. Hosmer-Lemeshow goodness-of-fit test indicated a satisfactory model fit with a p-value of 0.431. The calculator is available at: https://neurodx.shinyapps.io/WEBDEVICE/. CONCLUSION: The predictive calculator offers a substantial contribution to the clinical toolkit for estimating the likelihood of adequate intracranial aneurysm occlusion by WEB device embolization.
RESUMO
INTRODUCTION: The Woven EndoBridge (WEB) device is emerging as a novel therapy for intracranial aneurysms, but its use for off-label indications requires further study. Using machine learning, we aimed to develop predictive models for complete occlusion after off-label WEB treatment and to identify factors associated with occlusion outcomes. METHODS: This multicenter, retrospective study included 162 patients who underwent off-label WEB treatment for intracranial aneurysms. Baseline, morphological, and procedural variables were utilized to develop machine-learning models predicting complete occlusion. Model interpretation was performed to determine significant predictors. Ordinal regression was also performed with occlusion status as an ordinal outcome from better (Raymond Roy Occlusion Classification [RROC] grade 1) to worse (RROC grade 3) status. Odds ratios (OR) with 95 % confidence intervals (CI) were reported. RESULTS: The best performing model achieved an AUROC of 0.8 for predicting complete occlusion. Larger neck diameter and daughter sac were significant independent predictors of incomplete occlusion. On multivariable ordinal regression, higher RROC grades (OR 1.86, 95 % CI 1.25-2.82), larger neck diameter (OR 1.69, 95 % CI 1.09-2.65), and presence of daughter sacs (OR 2.26, 95 % CI 0.99-5.15) were associated with worse aneurysm occlusion after WEB treatment, independent of other factors. CONCLUSION: This study found that larger neck diameter and daughter sacs were associated with worse occlusion after WEB therapy for aneurysms. The machine learning approach identified anatomical factors related to occlusion outcomes that may help guide patient selection and monitoring with this technology. Further validation is needed.
Assuntos
Procedimentos Endovasculares , Aneurisma Intracraniano , Aprendizado de Máquina , Uso Off-Label , Humanos , Aneurisma Intracraniano/terapia , Aneurisma Intracraniano/diagnóstico por imagem , Masculino , Feminino , Estudos Retrospectivos , Pessoa de Meia-Idade , Resultado do Tratamento , Idoso , Fatores de Risco , Procedimentos Endovasculares/instrumentação , Procedimentos Endovasculares/efeitos adversos , Prótese Vascular , Desenho de Prótese , Técnicas de Apoio para a Decisão , Implante de Prótese Vascular/instrumentação , Implante de Prótese Vascular/efeitos adversos , Adulto , Tomada de Decisão Clínica , Medição de RiscoRESUMO
BACKGROUND: The comparative efficacy and safety of first-generation flow diverters (FDs), Pipeline Embolization Device (PED) (Medtronic, Irvine, California), Silk (Balt Extrusion, Montmorency, France), Flow Re-direction Endoluminal Device (FRED) (Microvention, Tustin, California), and Surpass Streamline (Stryker Neurovascular, Fremont, California), is not directly established and largely inferred. PURPOSE: This study aimed to compare the efficacy of different FDs in treating sidewall ICA intracranial aneurysms. METHODS: We conducted a retrospective review of prospectively maintained databases from eighteen academic institutions from 2009-2016, comprising 444 patients treated with one of four devices for sidewall ICA aneurysms. Data on demographics, aneurysm characteristics, treatment outcomes, and complications were analyzed. Angiographic and clinical outcomes were assessed using various imaging modalities and modified Rankin Scale (mRS). Propensity score weighting was employed to balance confounding variables. The data analysis used Kaplan-Meier curves, logistic regression, and Cox proportional-hazards regression. RESULTS: While there were no significant differences in retreatment rates, functional outcomes (mRS 0-1), and thromboembolic complications between the four devices, the probability of achieving adequate occlusion at the last follow-up was highest in Surpass device (HR: 4.59; CI: 2.75-7.66, pâ¯< 0.001), followed by FRED (HR: 2.23; CI: 1.44-3.46, pâ¯< 0.001), PED (HR: 1.72; CI: 1.10-2.70, pâ¯= 0.018), and Silk (HR: 1.0 ref. standard). The only hemorrhagic complications were with Surpass (1%). CONCLUSION: All the first-generation devices achieved good clinical outcomes and retreatment rates in treating ICA sidewall aneurysms. Prospective studies are needed to explore the nuanced differences between these devices in the long term.
RESUMO
BACKGROUND AND PURPOSE: Despite the numerous studies evaluating the occlusion rates of aneurysms following WEB embolization, there are limited studies identifying predictors of occlusion. Our purpose was to identify predictors of aneurysm occlusion and the need for retreatment. MATERIALS AND METHODS: This is a review of a prospectively maintained database across 30 academic institutions. We included patients with previously untreated cerebral aneurysms embolized using the WEB who had available intraprocedural data and long-term follow-up. RESULTS: We studied 763 patients with a mean age of 59.9 (SD, 11.7) years. Complete aneurysm occlusion was observed in 212/726 (29.2%) cases, and contrast stasis was observed in 485/537 (90.3%) of nonoccluded aneurysms. At the final follow-up, complete occlusion was achieved in 497/763 (65.1%) patients, and retreatment was required for 56/763 (7.3%) patients. On multivariable analysis, history of smoking, maximal aneurysm diameter, and the presence of an aneurysm wall branch were negative predictors of complete occlusion (OR, 0.5, 0.8, and 0.4, respectively). Maximal aneurysm diameter, the presence of an aneurysm wall branch, posterior circulation location, and male sex increase the chances of retreatment (OR, 1.2, 3.8, 3.0, and 2.3 respectively). Intraprocedural occlusion resulted in a 3-fold increase in the long-term occlusion rate and a 5-fold decrease in the retreatment rate (P < .001), offering a specificity of 87% and a positive predictive value of 85% for long-term occlusion. CONCLUSIONS: Intraprocedural occlusion can be used to predict the chance of long-term aneurysm occlusion and the need for retreatment after embolization with a WEB device. Smoking, aneurysm size, and the presence of an aneurysm wall branch are associated with decreased chances of successful treatment.
Assuntos
Embolização Terapêutica , Aneurisma Intracraniano , Humanos , Aneurisma Intracraniano/diagnóstico por imagem , Aneurisma Intracraniano/terapia , Aneurisma Intracraniano/cirurgia , Masculino , Feminino , Pessoa de Meia-Idade , Embolização Terapêutica/instrumentação , Embolização Terapêutica/métodos , Estudos Retrospectivos , Resultado do Tratamento , Idoso , Fatores de RiscoRESUMO
The Woven EndoBridge (WEB) device is primarily used for treating wide-neck intracranial bifurcation aneurysms under 10 mm. Limited data exists on its efficacy for large aneurysms. We aim to assess angiographic and clinical outcomes of the WEB device in treating large versus small aneurysms. We conducted a retrospective review of the WorldWide WEB Consortium database, from 2011 to 2022, across 30 academic institutions globally. Propensity score matching (PSM) was employed to compare small and large aneurysms on baseline characteristics. A total of 898 patients were included. There was no significant difference observed in clinical presentations, smoking status, pretreatment mRS, presence of multiple aneurysms, bifurcation location, or prior treatment between the two groups. After PSM, 302 matched pairs showed significantly lower last follow-up adequate occlusion rates (81% vs 90%, p = 0.006) and higher retreatment rates (12% vs 3.6%, p < 0.001) in the large aneurysm group. These findings may inform treatment decisions and patient counseling. Future studies are needed to further explore this area.
Assuntos
Procedimentos Endovasculares , Aneurisma Intracraniano , Pontuação de Propensão , Humanos , Aneurisma Intracraniano/terapia , Aneurisma Intracraniano/cirurgia , Masculino , Feminino , Pessoa de Meia-Idade , Estudos Retrospectivos , Idoso , Procedimentos Endovasculares/métodos , Resultado do Tratamento , Adulto , Embolização Terapêutica/métodosRESUMO
OBJECTIVE: This study was conducted to investigate the impact of antiplatelet administration in the periprocedural period on the occurrence of thromboembolic complications (TECs) in patients undergoing treatment using the Woven EndoBridge (WEB) device for intracranial wide-necked bifurcation aneurysms. The primary objective was to assess whether the use of antiplatelets in the pre- and postprocedural phases reduces the likelihood of developing TECs, considering various covariates. METHODS: A retrospective multicenter observational study was conducted within the WorldWideWEB Consortium and comprised 38 academic centers with endovascular treatment capabilities. Univariable and multivariable logistic regression analyses were performed to determine the association between antiplatelet use and TECs, adjusting for covariates. Missing predictor data were addressed using multiple imputation. RESULTS: The study comprised two cohorts: one addressing general thromboembolic events and consisting of 1412 patients, among whom 103 experienced TECs, and another focusing on symptomatic thromboembolic events and comprising 1395 patients, of whom 50 experienced symptomatic TECs. Preprocedural antiplatelet use was associated with a reduced likelihood of overall TECs (OR 0.32, 95% CI 0.19-0.53, p < 0.001) and symptomatic TECs (OR 0.49, 95% CI 0.25-0.95, p = 0.036), whereas postprocedural antiplatelet use showed no significant association with TECs. The study also revealed additional predictors of TECs, including stent use (overall: OR 4.96, 95% CI 2.38-10.3, p < 0.001; symptomatic: OR 3.24, 95% CI 1.26-8.36, p = 0.015), WEB single-layer sphere (SLS) type (overall: OR 0.18, 95% CI 0.04-0.74, p = 0.017), and posterior circulation aneurysm location (symptomatic: OR 18.43, 95% CI 1.48-230, p = 0.024). CONCLUSIONS: The findings of this study suggest that the preprocedural administration of antiplatelets is associated with a reduced likelihood of TECs in patients undergoing treatment with the WEB device for wide-necked bifurcation aneurysms. However, postprocedural antiplatelet use did not show a significant impact on TEC occurrence.
Assuntos
Procedimentos Endovasculares , Aneurisma Intracraniano , Inibidores da Agregação Plaquetária , Tromboembolia , Humanos , Inibidores da Agregação Plaquetária/administração & dosagem , Inibidores da Agregação Plaquetária/uso terapêutico , Masculino , Feminino , Estudos Retrospectivos , Pessoa de Meia-Idade , Tromboembolia/prevenção & controle , Tromboembolia/etiologia , Idoso , Procedimentos Endovasculares/métodos , Complicações Pós-Operatórias/prevenção & controle , Complicações Pós-Operatórias/etiologia , Complicações Pós-Operatórias/epidemiologia , AdultoRESUMO
BACKGROUND: The Woven EndoBridge (WEB) devices have been used for treating wide neck bifurcation aneurysms (WNBAs) with several generational enhancements to improve clinical outcomes. The original device dual-layer (WEB DL) was replaced by a single-layer (WEB SL) device in 2013. This study aimed to compare the effectiveness and safety of these devices in managing intracranial aneurysms. METHODS: A multicenter cohort study was conducted, and data from 1,289 patients with intracranial aneurysms treated with either the WEB SL or WEB DL devices were retrospectively analyzed. Propensity score matching was utilized to balance the baseline characteristics between the two groups. Outcomes assessed included immediate occlusion rate, complete occlusion at last follow-up, retreatment rate, device compaction, and aneurysmal rupture. RESULTS: Before propensity score matching, patients treated with the WEB SL had a significantly higher rate of complete occlusion at the last follow-up and a lower rate of retreatment. After matching, there was no significant difference in immediate occlusion rate, retreatment rate, or device compaction between the WEB SL and DL groups. However, the SL group maintained a higher rate of complete occlusion at the final follow-up. Regression analysis showed that SL was associated with higher rates of complete occlusion (OR: 0.19; CI: 0.04 to 0.8, p = 0.029) and lower rates of retreatment (OR: 0.12; CI: 0 to 4.12, p = 0.23). CONCLUSION: The WEB SL and DL devices demonstrated similar performances in immediate occlusion rates and retreatment requirements for intracranial aneurysms. The SL device showed a higher rate of complete occlusion at the final follow-up.
Assuntos
Embolização Terapêutica , Procedimentos Endovasculares , Aneurisma Intracraniano , Humanos , Resultado do Tratamento , Aneurisma Intracraniano/cirurgia , Aneurisma Intracraniano/etiologia , Embolização Terapêutica/efeitos adversos , Pontuação de Propensão , Estudos Retrospectivos , Estudos de Coortes , Procedimentos Endovasculares/efeitos adversosRESUMO
BACKGROUND: The Woven EndoBridge (WEB) device is frequently used for the treatment of intracranial aneurysms. Postoperative management, including the use of aspirin, varies among clinicians and institutions, but its impact on the outcomes of the WEB has not been thoroughly investigated. METHODS: This was a retrospective, multicenter study involving 30 academic institutions in North America, South America, and Europe. Data from 1492 patients treated with the WEB device were included. Patients were categorized into two groups based on their postoperative use of aspirin (aspirin group: n=1124, non-aspirin group: n=368). Data points included patient demographics, aneurysm characteristics, procedural details, complications, and angiographic and functional outcomes. Propensity score matching (PSM) was applied to balance variables between the two groups. RESULTS: Prior to PSM, the aspirin group exhibited significantly higher rates of modified Rankin scale (mRS) mRS 0-1 and mRS 0-2 (89.8% vs 73.4% and 94.1% vs 79.8%, p<0.001), lower rates of mortality (1.6% vs 8.6%, p<0.001), and higher major compaction rates (13.4% vs 7%, p<0.001). Post-PSM, the aspirin group showed significantly higher rates of retreatment (p=0.026) and major compaction (p=0.037) while maintaining its higher rates of good functional outcomes and lower mortality rates. In the multivariable regression, aspirin was associated with higher rates of mRS 0-1 (OR 2.166; 95% CI 1.16 to 4, p=0.016) and mRS 0-2 (OR 2.817; 95% CI 1.36 to 5.88, p=0.005) and lower rates of mortality (OR 0.228; 95% CI 0.06 to 0.83, p=0.025). However, it was associated with higher rates of retreatment (OR 2.471; 95% CI 1.11 to 5.51, p=0.027). CONCLUSIONS: Aspirin use post-WEB treatment may lead to better functional outcomes and lower mortality but with higher retreatment rates. These insights are crucial for postoperative management after WEB procedures, but further studies are necessary for validation.
RESUMO
OBJECTIVE: The Woven EndoBridge (WEB) device is an intrasaccular flow disruptor designed for wide-necked bifurcation aneurysms. These aneurysms may require the use of a concomitant stent. The objective of this study was to determine the clinical and radiological outcomes of patients undergoing stent-assisted WEB treatment. In addition, the authors also sought to determine the predictors of a concomitant stent in aneurysms treated with the WEB device. METHODS: The data for this study were taken from the WorldWideWEB Consortium, an international multicenter cohort including patients treated with the WEB device. Aneurysms were classified into two groups based on treatment: stent-assisted WEB and WEB device alone. The authors compared clinical and radiological outcomes of both groups. Univariable and multivariable binary logistic regression analyses were performed to determine factors that predispose to stent use. RESULTS: The study included 691 intracranial aneurysms (31 with stents and 660 without stents) treated with the WEB device. The adequate occlusion status did not differ between the two groups at the latest follow-up (83.3% vs 85.6%, p = 0.915). Patients who underwent stenting had more thromboembolic (32.3% vs 6.5%, p < 0.001) and procedural (16.1% vs 3.0%, p < 0.001) complications. Aneurysms treated with a concomitant stent had wider necks, greater heights, and lower dome-to-neck ratios. Increasing neck size was the only significant predictor for stent use. CONCLUSIONS: This study demonstrates that there is no difference in the degree of aneurysm occlusion between the two groups; however, complications were more frequent in the stent group. In addition, a wider aneurysm neck predisposes to stent assistance in WEB-treated aneurysms.
Assuntos
Embolização Terapêutica , Procedimentos Endovasculares , Aneurisma Intracraniano , Humanos , Aneurisma Intracraniano/diagnóstico por imagem , Aneurisma Intracraniano/cirurgia , Resultado do Tratamento , Estudos Retrospectivos , StentsRESUMO
BACKGROUND: The Woven EndoBridge (WEB) device is a novel intrasaccular flow disruptor tailored for bifurcation aneurysms. We aim to describe the degree of aneurysm occlusion at the latest follow-up, and the rate of complications of aneurysms treated with the WEB device stratified according to rupture status. METHODS: Our data were taken from the WorldWideWeb Consortium, an international multicenter cohort including patients treated with the WEB device. Aneurysms were classified into two groups: ruptured and unruptured. We compared clinical and radiologic outcomes of both groups. Propensity score matching (PSM) was done to match according to age, gender, bifurcation, location, prior treatment, neck, height, dome width, daughter sac, incorporated branch, pretreatment antiplatelets, and last imaging follow-up. RESULTS: The study included 676 patients with 691 intracranial aneurysms (529 unruptured and 162 ruptured) treated with the WEB device. The PSM analysis had 55 pairs. In both the unmatched (85.8% vs 84.3%, p=0.692) and matched (94.4% vs 83.3%, p=0.066) cohorts there was no significant difference in the adequate occlusion rate at the last follow-up. Likewise, there were no significant differences in both ischemic and hemorrhagic complications between the two groups. There was no documented aneurysm rebleeding after WEB device implantation. CONCLUSION: There was no significant difference in both the radiologic outcomes and complications between unruptured and ruptured aneurysms. Our findings support the feasibility of treatment of ruptured aneurysms with the WEB device.
Assuntos
Aneurisma Roto , Embolização Terapêutica , Procedimentos Endovasculares , Aneurisma Intracraniano , Humanos , Resultado do Tratamento , Procedimentos Endovasculares/métodos , Estudos Retrospectivos , Aneurisma Roto/diagnóstico por imagem , Aneurisma Roto/cirurgia , Aneurisma Intracraniano/diagnóstico por imagem , Aneurisma Intracraniano/cirurgia , Embolização Terapêutica/métodosRESUMO
Several studies have shown promising outcomes of the Woven EndoBridge (WEB) device for the treatment of wide-necked intracranial bifurcation aneurysms. This is a multicenter study attempts to explore the changes in trends and treatment outcomes over time for WEB embolization of intracranial aneurysms. The WorldWideWEB consortium is a retrospective multicenter collaboration of data from international centers spanning from January 2011 and June 2021, with no limitations on aneurysm location or rupture status. Both bifurcation and sidewall aneurysms were included. These patients were stratified based on treatment year into five treatment intervals: 2011-2015 (N = 66), 2016-2017 (N = 77), 2018 (N = 66), 2019 (N = 300), and 2020-2021 (N = 173). Patient characteristics and angiographic and clinical outcomes were compared between these time intervals. This study comprised 671 patients (median age 61.4 years; 71.2% female) with 682 intracranial aneurysms. Over time, we observed an increasing tendency to treat patients presenting with ruptured aneurysms and aneurysms with smaller neck, diameter, and dome widths. Furthermore, we observed a trend towards more off-label use of the WEB for sidewall aneurysms and increased adoption of transradial access for WEB deployment. Moreover, the proportion of patients with adequate WEB occlusion immediately and at last follow-up was significantly higher in more recent year cohorts, as well as lower rates of compaction and retreatment. Mortality and complications did not differ over time. This learning curve study suggests improved experience using the WEB for the treatment of intracranial aneurysms and has yielded higher rates of adequate occlusion over time.
Assuntos
Embolização Terapêutica , Procedimentos Endovasculares , Aneurisma Intracraniano , Humanos , Feminino , Pessoa de Meia-Idade , Masculino , Aneurisma Intracraniano/diagnóstico por imagem , Aneurisma Intracraniano/cirurgia , Resultado do Tratamento , Estudos RetrospectivosRESUMO
BACKGROUND: The Woven EndoBridge (WEB) device has Food and Drug Administration approval for treatment of wide-necked intracranial bifurcation aneurysms. The WEB device has been shown to result in adequate occlusion in bifurcation aneurysms overall, but its usefulness in the individual bifurcation locations has been evaluated separately only in few case series, which were limited by small sample sizes. OBJECTIVE: To compare angiographic and clinical outcomes after treatment of bifurcation aneurysms at various locations, including anterior communicating artery (AComA), anterior cerebral artery (ACA) bifurcation distal to AComA, basilar tip, internal carotid artery (ICA) bifurcation, and middle cerebral artery (MCA) bifurcation aneurysms using the WEB device. METHODS: A retrospective cohort analysis was conducted at 22 academic institutions worldwide to compare treatment outcomes of patients with intracranial bifurcation aneurysms using the WEB device. Data include patient and aneurysm characteristics, procedural details, angiographic and functional outcomes, and complications. RESULTS: A total of 572 aneurysms were included. MCA (36%), AComA (35.7%), and basilar tip (18.9%) aneurysms were most common. The rate of adequate aneurysm occlusion was significantly higher for basilar tip (91.6%) and ICA bifurcation (96.7%) aneurysms and lower for ACA bifurcation (71.4%) and AComA (80.6%) aneurysms (p=0.04). CONCLUSION: To our knowledge, this is the most extensive study to date that compares the treatment of different intracranial bifurcation aneurysms using the WEB device. Basilar tip and ICA bifurcation aneurysms showed significantly higher rates of aneurysm occlusion than other locations.
Assuntos
Doenças das Artérias Carótidas , Embolização Terapêutica , Procedimentos Endovasculares , Aneurisma Intracraniano , Humanos , Estudos Retrospectivos , Resultado do Tratamento , Aneurisma Intracraniano/diagnóstico por imagem , Aneurisma Intracraniano/terapia , Doenças das Artérias Carótidas/terapiaRESUMO
Background The Woven EndoBridge (WEB) device was explicitly designed for wide-neck intracranial bifurcation aneurysms. Small-scale reports have evaluated the off-label use of WEB devices for the treatment of sidewall aneurysms, with promising outcomes. Purpose To compare the angiographic and clinical outcomes of the WEB device for the treatment of sidewall aneurysms compared with the treatment of bifurcation aneurysms. Materials and Methods A retrospective review of the WorldWideWEB Consortium, a synthesis of retrospective databases spanning from January 2011 to June 2021 at 22 academic institutions in North America, South America, and Europe, was performed to identify patients with intracranial aneurysms treated with the WEB device. Characteristics and outcomes were compared between bifurcation and sidewall aneurysms. Propensity score matching (PSM) was used to match by age, pretreatment ordinal modified Rankin Scale score, ruptured aneurysms, location of aneurysm, multiple aneurysms, prior treatment, neck, height, dome width, daughter sac, and incorporated branch. Results A total of 683 intracranial aneurysms were treated using the WEB device in 671 patients (median age, 61 years [IQR, 53-68 years]; male-to-female ratio, 1:2.5). Of those, 572 were bifurcation aneurysms and 111 were sidewall aneurysms. PSM was performed, resulting in 91 bifurcation and sidewall aneurysms pairs. No significant difference was observed in occlusion status at last follow-up, deployment success, or complication rates between the two groups. Conclusion No significantly different outcomes were observed following the off-label use of the Woven EndoBridge, or WEB, device for treatment of sidewall aneurysms compared with bifurcation aneurysms. The correct characterization of the sidewall aneurysm location, neck angle, and size is crucial for successful treatment and lower retreatment rate. © RSNA, 2022 See also the editorial by Hetts in this issue.
Assuntos
Aneurisma Roto , Embolização Terapêutica , Procedimentos Endovasculares , Aneurisma Intracraniano , Aneurisma Roto/terapia , Embolização Terapêutica/métodos , Procedimentos Endovasculares/métodos , Feminino , Humanos , Aneurisma Intracraniano/diagnóstico por imagem , Aneurisma Intracraniano/cirurgia , Masculino , Pessoa de Meia-Idade , Estudos Retrospectivos , Resultado do TratamentoRESUMO
OBJECTIVE: Transradial access (TRA) is commonly utilized in neurointerventional procedures. This study compared the technical and clinical outcomes of the use of TRA versus those of transfemoral access (TFA) for intracranial aneurysm embolization with the Woven EndoBridge (WEB) device. METHODS: This is a secondary analysis of the Worldwide WEB Consortium, which comprises multicenter data related to adult patients with intracranial aneurysms who were managed with the WEB device. These aneurysms were categorized into two groups: those who were treated with TRA or TFA. Patient and aneurysm characteristics and technical and clinical outcomes were compared between groups. Propensity score matching (PSM) was used to match groups according to the following baseline characteristics: age, sex, subarachnoid hemorrhage, aneurysm location, bifurcation aneurysm, aneurysm with incorporated branch, neck width, aspect ratio, dome width, and elapsed time since the last follow-up imaging evaluation. RESULTS: This study included 682 intracranial aneurysms (median [interquartile range] age 61.3 [53.0-68.0] years), of which 561 were treated with TFA and 121 with TRA. PSM resulted in 65 matched pairs. After PSM, both groups had similar characteristics, angiographic and functional outcomes, and rates of retreatment, thromboembolic and hemorrhagic complications, and death. TFA was associated with longer procedure length (median 96.5 minutes vs 72.0 minutes, p = 0.006) and fluoroscopy time (28.2 minutes vs 24.8 minutes, p = 0.037) as compared with TRA. On the other hand, deployment issues were more common in those treated with TRA, but none resulted in permanent complications. CONCLUSIONS: TRA has comparable outcomes, with shorter procedure and fluoroscopy time, to TFA for aneurysm embolization with the WEB device.
Assuntos
Procedimentos Endovasculares/métodos , Aneurisma Intracraniano/cirurgia , Stents , Adulto , Idoso , Aneurisma Roto/prevenção & controle , Aneurisma Roto/terapia , Angiografia Digital , Angiografia por Tomografia Computadorizada , Bases de Dados Factuais , Embolização Terapêutica , Feminino , Seguimentos , Humanos , Aneurisma Intracraniano/diagnóstico por imagem , Masculino , Pessoa de Meia-Idade , Estudos Retrospectivos , Resultado do TratamentoRESUMO
BACKGROUND: Data regarding the safety and efficacy of flow diverting stents (FDS) in the treatment of middle cerebral artery (MCA) bifurcation aneurysms are scarce and limited to small single center series, with particular concern for increased risk of ischemic complications with jailing one of the M2 branches. METHODS: Prospectively-maintained databases at six North American and European centers were queried for patients harboring MCA bifurcation aneurysms undergoing treatment with FDS (2011-2018). The pertinent clinical and radiographic data were collected and analyzed. RESULTS: 87 patients (median age 60 years, 69% females) harboring 87 aneurysms were included. The majority of aneurysms were unruptured (79%); 75.9% were saccular with a median maximal diameter of 8.5 mm. Radiographic imaging follow-up was available in 88.5% of cases at a median of 16.3 months post-treatment, showing complete occlusion in 59% and near complete occlusion (90-99%) in 18% of aneurysms. The overall rate of ischemic and hemorrhagic complications was 8% and 1.1%, respectively. Symptomatic and permanent complications were encountered in 5.7% and 2.3% of patients respectively, with retreatment pursued in 2.3% of patients. Jailed branch occlusion was detected in 11.5% of cases, with clinical sequelae in 2.3%. Last follow-up modified Rankin Scale of 0-2 was noted in 96.8% of patients. On multivariate analysis, male sex was the only independent predictor of aneurysmal persistence at last follow-up imaging (p=0.019). CONCLUSION: FDS treatment for MCA bifurcation aneurysms is feasible, with comparable safety and efficacy profiles to other available endovascular options when utilized in carefully selected aneurysms. Jailing of M2 branches was not associated with a higher risk of post-procedural ischemic complications.