Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros

Base de dados
Ano de publicação
Tipo de documento
Intervalo de ano de publicação
1.
Appl Radiat Isot ; 70(9): 2006-11, 2012 Sep.
Artigo em Inglês | MEDLINE | ID: mdl-22421395

RESUMO

The radionuclide (68)Ga is one of the few positron emitters that can be prepared in-house without the use of a cyclotron. It disintegrates to the ground state of (68)Zn partially by positron emission (89.1%) with a maximum energy of 1899.1 keV, and partially by electron capture (10.9%). This nuclide has been standardized in the frame of a cooperation project between the Radionuclide Metrology laboratories from CIEMAT (Spain) and CNEA (Argentina). Measurements involved several techniques: 4πß-γ coincidences, integral gamma counting and Liquid Scintillation Counting using the triple to double coincidence ratio and the CIEMAT/NIST methods. Given the short half-life of the radionuclide assayed, a direct comparison between results from both laboratories was excluded and a comparison of experimental efficiencies of similar NaI detectors was used instead.


Assuntos
Radioisótopos de Gálio/química , Radioisótopos de Gálio/normas , Contagem de Cintilação/instrumentação , Contagem de Cintilação/normas , Desenho de Equipamento , Análise de Falha de Equipamento , Radioisótopos de Gálio/análise , Meia-Vida , Internacionalidade , Doses de Radiação , Padrões de Referência , Valores de Referência
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA