1.
World J Methodol ; 13(5): 502-509, 2023 Dec 20.
Artigo em Inglês | MEDLINE | ID: mdl-38229936

RESUMO

BACKGROUND: The ExeterTM Universal cemented femoral component is widely used for total hip replacement surgery. Although there have been few reports of femoral component fracture, removal of a broken femoral stem can be a challenging procedure. CASE SUMMARY: A 54-year-old man with a Dorr A femur sustained a refracture of a primary ExeterTM stem, two years after receiving a revision using a cement-within-cement technique (CWC) through an extended trochanteric osteotomy (ETO). The technical problems related to the CWC technique and the ETO played a major role in the stem fatigue refracture. We performed revision surgery and removed the distal cement using a cortical femoral window technique, followed by re-implantation with an uncemented, modular, distally-fixed uncemented stem. The patient experienced an uneventful postoperative recovery. CONCLUSION: Re-fracture of a modern femoral ExeterTM stem is a rare event, but technical complications related to revision surgery can lead to this outcome. The cortical window osteotomy technique can facilitate the removal of a broken stem and cement, allowing for prosthetic re-implantation under direct vision and avoiding ETO-related complications.