1.
Appl Radiat Isot ; 83 Pt C: 222-6, 2014 Jan.
Artigo em Inglês | MEDLINE | ID: mdl-23810727

RESUMO

MOSkin detectors were studied to perform real-time in vivo dose measurements in high dose rate prostate brachytherapy. Measurements were performed inside an urethral catheter in a gel phantom simulating a real prostate implant. Measured and expected doses were compared and the discrepancy was found to be within 8.9% and 3.8% for single MOSkin and dual-MOSkin configurations, respectively. Results show that dual-MOSkin detectors can be profitably adopted in prostate brachytherapy treatments to perform real-time in vivo dosimetry inside the urethra.


Assuntos
Braquiterapia , Imagens de Fantasmas , Neoplasias da Próstata/radioterapia , Dosagem Radioterapêutica , Estudos de Viabilidade , Humanos , Masculino