Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 2 de 2
Filtrar
Mais filtros

Base de dados
Ano de publicação
Tipo de documento
Intervalo de ano de publicação
2.
Rev Sci Instrum ; 92(3): 033516, 2021 Mar 01.
Artigo em Inglês | MEDLINE | ID: mdl-33820046

RESUMO

Electron-positron pairs, produced in intense laser-solid interactions, are diagnosed using magnetic spectrometers with image plates, such as the National Ignition Facility Electron-Positron-Proton Spectrometers (EPPSs). Although modeling can help infer the quantitative value, the accuracy of the models needs to be verified to ensure measurement quality. The dispersion of low-energy electrons and positrons may be affected by fringe magnetic fields near the entrance of the EPPS. We have calibrated the EPPS with six electron beams from a Siemens Oncor linear accelerator (linac) ranging in energy from 2.7 MeV to 15.2 MeV as they enter the spectrometer. A Geant4 traveling-wave optical parametric amplifier of superfluorescence Monte Carlo simulation was set up to match depth dose curves and lateral profiles measured in water at 100 cm source-surface distance. An accurate relationship was established between the bending magnet current setting and the energy of the electron beam at the exit window. The simulations and measurements were used to determine the energy distributions of the six electron beams at the EPPS slit. Analysis of the scanned image plates together with the determined energy distribution arriving in the spectrometer provides improved dispersion curves for the EPPS.

SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA