Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 3 de 3
Filtrar
Mais filtros

Base de dados
Ano de publicação
Tipo de documento
Intervalo de ano de publicação
2.
JACC Case Rep ; 29(3): 102192, 2024 Feb 07.
Artigo em Inglês | MEDLINE | ID: mdl-38361571

RESUMO

Misloading during transcatheter aortic valve replacement (TAVR) is rare but can cause unpredictable valve release if unrecognized. We describe how to identify a misloaded ACURATE neo2 device, and 3 methods to solve this by using a modified technique of valve deployment, ipsilateral extraction, and contralateral valve externalization with extracorporeal valve release.

3.
Int J Cardiol ; 377: 26-32, 2023 04 15.
Artigo em Inglês | MEDLINE | ID: mdl-36640966

RESUMO

BACKGROUND: Cardiac tamponade caused by temporary right ventricular (RV) pacemaker perforation is a rare but serious complication in transcatheter aortic valve replacement (TAVR). AIMS: To study the incidence of temporary pacemaker related cardiac tamponade in TAVR, and the relation to the type of pacemaker lead used in periprocedural temporary transvenous pacing. METHODS: A single center registry of transfemoral TAVRs in 2014-2020. Main inclusion criterion was peri-operative use of a temporary RV pacing lead. Main exclusion criteria were a preoperatively implanted permanent pacemaker or the exclusive use of left ventricular guidewire pacing. Incident cardiac tamponade was classified as pacemaker lead related, or other. Patients were grouped according to type of temporary RV pacing wire. RESULTS: 810 patients were included (age 80.5 ± 7.3 [mean ± standard deviation], female 319, 39.4%). Of these, 566 (69.9%) received a standard RV temporary pacing wire (RV-TPW), and 244 (30.1%) received temporary RV pacing through a permanent, passive pacemaker lead (RV-TPPL). In total, 18 (2.2%) events of cardiac tamponade occurred, 12 (67%) were pacemaker lead related. All pacemaker lead-related cardiac tamponades occurred in the group who received a standard RV-TPW and none in the group who received RV-TPPL (n = 12 [2.1%] vs. n = 0 [0%], p = 0.022). No difference in cardiac tamponade due to other causes was seen between the groups (p = 0.82). CONCLUSIONS: The use of soft-tip RV-TPPL was associated with a lower risk of pacemaker related cardiac tamponade in TAVR. When perioperative pacing is indicated, temporary RV-TPPL may contribute to a significant reduction of cardiac tamponade in TAVR.


Assuntos
Estenose da Valva Aórtica , Tamponamento Cardíaco , Marca-Passo Artificial , Substituição da Valva Aórtica Transcateter , Humanos , Feminino , Idoso , Idoso de 80 Anos ou mais , Substituição da Valva Aórtica Transcateter/efeitos adversos , Tamponamento Cardíaco/diagnóstico , Tamponamento Cardíaco/epidemiologia , Tamponamento Cardíaco/etiologia , Resultado do Tratamento , Marca-Passo Artificial/efeitos adversos , Estimulação Cardíaca Artificial , Estenose da Valva Aórtica/cirurgia , Fatores de Risco , Valva Aórtica/cirurgia
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA