1.
Micromachines (Basel) ; 15(7)2024 Jul 17.
Artigo em Inglês | MEDLINE | ID: mdl-39064432

RESUMO

Thin-film electrode arrays (TFEAs) have been developed as an alternative to conventional electrode arrays (CEAs) used in cochlear implants. However, TFEAs produced by microfabrication techniques have not yet been used clinically because their structural and mechanical properties are far from those of CEAs. The aim of this study is to design, fabricate, and investigate the mechanical and tribological behavior and evaluate the performance of different TFEA designs. Finite Element Analysis (FEA) is performed to determine the elastic properties of several designs. A custom-build experimental setup is designed to observe the tribological behavior in different speeds and environments where frictional (lateral) and vertical force (normal force) are measured on a flat surface and within artificial cochlea. According to the FEA results, the maximum stiffness of the CEA is 37.93 mN/mm and 0.363 mN/mm and TFEA-4 has a maximum stiffness of 39.08 mN/mm and 0.306 mN/mm in the longitudinal and transverse axes, respectively. It is shown experimentally that adding a dummy wire to the carrier of the EA enhances both its longitudinal and transverse stiffness, thereby postponing the initiation of dynamic sliding due to the elevated buckling limit. It is also revealed that the type of TFEA support structure affects both normal and frictional forces, as well as the coefficient of friction.

SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA