1.
Appl Radiat Isot
; 126: 296-299, 2017 Aug.
Artigo
em Inglês
| MEDLINE
| ID: mdl-28109700
RESUMO
A measurement station dedicated for quantitative radiological characterisation of naturally occurring radionuclides in a metallurgical company and based on gamma-ray spectrometry was developed. The station is intended for performing quality control of final non-ferrous metal products and for radiological checks of incoming materials. A low-background point-contact HPGe-detector was used and the signal was split in two branches to enable collecting simultaneously spectra with high amplification (for gamma-ray energies below 250keV) and low amplification.