Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros

Base de dados
Ano de publicação
Tipo de documento
Intervalo de ano de publicação
1.
Catheter Cardiovasc Interv ; 52(1): 67-72, 2001 Jan.
Artigo em Inglês | MEDLINE | ID: mdl-11146526

RESUMO

This article reports the status of a new cardiovascular fluoroscopy benchmarking phantom. A joint working group of the Society for Cardiac Angiography and Interventions (SCA&I) and the National Electrical Manufacturers Association (NEMA) developed the phantom. The device was adopted as NEMA standard XR 21-2000, "Characteristics of and Test Procedures for a Phantom to Benchmark Cardiac Fluoroscopic and Photographic Performance," in August 2000. The test ensemble includes imaging field geometry, spatial resolution, low-contrast iodine detectability, working thickness range, visibility of moving targets, and phantom entrance dose. The phantom tests systems under conditions simulating normal clinical use for fluoroscopically guided invasive and interventional procedures. Test procedures rely on trained human observers.


Assuntos
Cateterismo Cardíaco/instrumentação , Fluoroscopia/instrumentação , Fluoroscopia/normas , Imagens de Fantasmas , Benchmarking , Cateterismo Cardíaco/métodos , Desenho de Equipamento , Segurança de Equipamentos , Humanos , Sensibilidade e Especificidade
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA