RESUMO
The dosimetric output of a 6FFF beam, produced from a Varian TrueBeam linac exhibited an unexpected downward trend over time that was contrary to well-established expectations. To elucidate the cause of this uncharacteristic trend, a review of the linac's quality control results over its lifetime was performed, including, constancy checks of the dosimetric output, beam energy, flatness and symmetry, and percentage depth dose characteristics. These results were supplemented with a comprehensive series of measurements including flatness and symmetry measurements with a 1D-diode array, high-resolution measurements of the photon beam's build-up region with a parallel-plate chamber and measurement of the beam's output as a function of the x-ray target position. The review of the linac's QC results and supplemental tests identified no deviations in the linac's performance from its commissioning and baseline measurements. However, the 6FFF beam output exhibited a significant dependence on the target location relative to its default position, increasing by 5.43 % with a 0.5 mm target translation, indicating that target degradation was the cause of the atypical output trend. The change in output behaviour was believed to be the result of primary electrons escaping the degraded target and interacting with the linac's monitor chamber. Replacement of the x-ray target caused the 6FFF output to realign with expected trends. Target degradation was uncovered due to a robust quality control trending database and awareness of typical output behaviour. These results demonstrate the importance of data trending to identify component failure and provide centres with knowledge to recognise this potential fault.