Atrial therapies reduce atrial arrhythmia burden in defibrillator patients.
Circulation
; 104(9): 1023-8, 2001 Aug 28.
Article
em En
| MEDLINE
| ID: mdl-11524396
ABSTRACT
BACKGROUND:
Approximately 25% of patients who receive an implantable cardioverter-defibrillator (ICD) to treat ventricular tachyarrhythmias have documented atrial tachyarrhythmias before implantation. This study assessed the ability of device-based prevention and termination therapies to reduce the burden of spontaneous atrial tachyarrhythmias. METHODS ANDRESULTS:
Patients with a standard indication for the implantation of an ICD and 2 episodes of atrial tachyarrhythmias in the preceding year received a dual-chamber ICD (Medtronic 7250 Jewel AF) that uses pacing and shock therapies for prevention and/or termination of atrial tachyarrhythmias. In a multicenter trial, patients were randomized to 3-month periods with atrial therapies "on" or "off" and subsequently crossed over. Analysis was performed on the 52 of 269 patients who had episodes of atrial tachyarrhythmia and had >/=30 days of follow-up with atrial therapies on and off. The atrial therapies resulted in a reduction of atrial tachyarrhythmia burden from a mean of 58.5 to 7.8 h/mo. A paired analysis (Wilcoxon signed-rank test) showed that the median difference in burden (1.1 h/mo) was highly significant (P=0.007). When the subgroup of 41 patients treated only with atrial pacing therapies was analyzed, the reduction in burden persisted (P=0.01).CONCLUSIONS:
In this study, patients with a standard ICD indication and atrial tachyarrhythmias had a significant reduction in atrial tachyarrhythmia burden with use of atrial pacing and shock therapies.
Buscar no Google
Base de dados:
MEDLINE
Assunto principal:
Fibrilação Atrial
/
Taquicardia Ventricular
/
Desfibriladores Implantáveis
Idioma:
En
Ano de publicação:
2001
Tipo de documento:
Article