Transvenous transjugular renal core biopsy with a redesigned biopsy set including a blunt-tipped needle.
Sofocleous, Constantinos T; Bahramipour, Philip; Mele, Christopher; Hinrichs, Clay R; Barone, Alison; Abujudeh, Hani.
Afiliação
  • Sofocleous CT; Section of Vascular & Interventional, Department of Radiology, New Jersey Medical School, University Hospital, C320, 150 Bergen Street, Newark, NJ 07103-2406, USA.
Cardiovasc Intervent Radiol ; 25(2): 155-7, 2002.
Article em En | MEDLINE | ID: mdl-11901438
ABSTRACT
A novel 19-gauge, blunt-tipped, side cutting single throw, 70-cm long transjugular needle, specifically designed for transvenous kidney biopsy, was used in seven patients with high risk for bleeding. A mean of 4 device-passes (3-6) per patient resulted in a satisfactory specimen for pathological diagnosis. Immediate post-biopsy nonenhanced CT was performed to evaluate for bleeding at the biopsy site. All patients were observed for 2 hr after the procedure. No clinically significant immediate or late complication was noted.
Assuntos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Biópsia por Agulha / Rim Idioma: En Ano de publicação: 2002 Tipo de documento: Article