Transvenous transjugular renal core biopsy with a redesigned biopsy set including a blunt-tipped needle.
Cardiovasc Intervent Radiol
; 25(2): 155-7, 2002.
Article
em En
| MEDLINE
| ID: mdl-11901438
ABSTRACT
A novel 19-gauge, blunt-tipped, side cutting single throw, 70-cm long transjugular needle, specifically designed for transvenous kidney biopsy, was used in seven patients with high risk for bleeding. A mean of 4 device-passes (3-6) per patient resulted in a satisfactory specimen for pathological diagnosis. Immediate post-biopsy nonenhanced CT was performed to evaluate for bleeding at the biopsy site. All patients were observed for 2 hr after the procedure. No clinically significant immediate or late complication was noted.
Buscar no Google
Base de dados:
MEDLINE
Assunto principal:
Biópsia por Agulha
/
Rim
Idioma:
En
Ano de publicação:
2002