Your browser doesn't support javascript.
loading
Comparison between microscopic and automated differential leukocyte counts in the silver fox (Vulpes vulpes) and the blue fox (Alopex lagopus).
Moe, Randi Oppermann; Bakken, Morten; Brun-Hansen, Hege; Adnøy, Tormod; Morberg, Hanne.
Afiliação
  • Moe RO; Norwegian College of Veterinary Medicine, Oslo, Norway. Randi.O.Moe@online.no
Vet Clin Pathol ; 28(2): 65-70, 1999.
Article em En | MEDLINE | ID: mdl-12075528
Differential leukocyte (WBC) counts in blood from clinically healthy silver foxes (n=32) and blue foxes (n=37) obtained from an automated hematology analyzer (Technicon H*1 Hematology System) with canine software were compared with microscopic differential WBC counts (M-diff). There was good agreement between the automated differential cell count (A-diff) and the M-diff for neutrophil and lymphocyte percentages. The correlation was lower for monocyte percentages and variable for eosinophil percentages. There was no significant difference between the A-diff and M-diff in either fox species. The A-diff counts were very precise, and may be a good alternative to the traditional M-diff for screening populations of clinically healthy foxes or for studies on stress and animal welfare. Intercept values, however, indicated a constant bias that must be taken into account before interpreting results based on different methods of analysis
Buscar no Google
Base de dados: MEDLINE Idioma: En Ano de publicação: 1999 Tipo de documento: Article
Buscar no Google
Base de dados: MEDLINE Idioma: En Ano de publicação: 1999 Tipo de documento: Article