Your browser doesn't support javascript.
loading
Investigation of secondary neutron dose for 18 MV dynamic MLC IMRT delivery.
Howell, Rebecca M; Ferenci, Michele S; Hertel, Nolan E; Fullerton, Gary D.
Afiliação
  • Howell RM; Department of Radiation Oncology, Emory University School of Medicine, 1365 Clifton Road AT200, Atlanta, Georgia 30322, USA. rebecca@radonc.emory.org
Med Phys ; 32(3): 786-93, 2005 Mar.
Article em En | MEDLINE | ID: mdl-15839351
ABSTRACT
Secondary neutron doses from the delivery of 18 MV conventional and intensity modulated radiation therapy (IMRT) treatment plans were compared. IMRT was delivered using dynamic multileaf collimation (MLC). Additional measurements were made with static MLC using a primary collimated field size of 10 x 10 cm2 and MLC field sizes of 0 x 0, 5 x 5, and 10 x 10 cm2. Neutron spectra were measured and effective doses calculated. The IMRT treatment resulted in a higher neutron fluence and higher dose equivalent. These increases were approximately the ratio of the monitor units. The static MLC measurements were compared to Monte Carlo calculations. The actual component dimensions and materials for the Varian Clinac 2100/2300C including the MLC were modeled with MCNPX to compute the neutron fluence due to neutron production in and around the treatment head. There is excellent agreement between the calculated and measured neutron fluence for the collimated field size of 10 x 10 cm2 with the 0 x 0 cm2 MLC field. Most of the neutrons at the detector location for this geometry are directly from the accelerator head with a small contribution from room scatter. Future studies are needed to investigate the effect of different beam energies used in IMRT incorporating the effects of scattered photon dose as well as secondary neutron dose.
Assuntos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Neoplasias da Próstata / Radiometria / Planejamento da Radioterapia Assistida por Computador / Radioterapia Conformacional / Nêutrons Idioma: En Ano de publicação: 2005 Tipo de documento: Article
Buscar no Google
Base de dados: MEDLINE Assunto principal: Neoplasias da Próstata / Radiometria / Planejamento da Radioterapia Assistida por Computador / Radioterapia Conformacional / Nêutrons Idioma: En Ano de publicação: 2005 Tipo de documento: Article