Development of a 35-MHz piezo-composite ultrasound array for medical imaging.
Cannata, Jonathan M; Williams, Jay A; Zhou, Qifa; Ritter, Timothy A; Shung, K Kirk.
Afiliação
  • Cannata JM; NIH Resource on Medical Ultrasonic Transducer Technology, Department of Biomedical Engineering, University of Southern California, Los Angeles, CA, USA. cannata@usc.edu
Article em En | MEDLINE | ID: mdl-16471449
ABSTRACT
This paper discusses the development of a 64-element 35-MHz composite ultrasonic array. This array was designed primarily for ocular imaging applications, and features 2-2 composite elements mechanically diced out of a fine-grain high-density Navy Type VI ceramic. Array elements were spaced at a 50-micron pitch, interconnected via a custom flexible circuit and matched to the 50-ohm system electronics via a 75-ohm transmission line coaxial cable. Elevation focusing was achieved using a cylindrically shaped epoxy lens. One functional 64-element array was fabricated and tested. Bandwidths averaging 55%, 23-dB insertion loss, and crosstalk less than -24 dB were measured. An image of a tungsten wire target phantom was acquired using a synthetic aperture reconstruction algorithm. The results from this imaging test demonstrate resolution exceeding 50 microm axially and 100 microm laterally.
Assuntos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Transdutores / Aumento da Imagem / Cerâmica / Ultrassonografia / Desenho Assistido por Computador / Microeletrodos / Modelos Teóricos Idioma: En Ano de publicação: 2006 Tipo de documento: Article
Buscar no Google
Base de dados: MEDLINE Assunto principal: Transdutores / Aumento da Imagem / Cerâmica / Ultrassonografia / Desenho Assistido por Computador / Microeletrodos / Modelos Teóricos Idioma: En Ano de publicação: 2006 Tipo de documento: Article