Development of a high-performance liquid chromatography method for the determination of caspofungin with amperometric detection and its application to in vitro microdialysis experiments.
Traunmüller, Friederike; Steiner, Ilka; Zeitlinger, Markus; Joukhadar, Christian.
Afiliação
  • Traunmüller F; Department of Clinical Pharmacology, Division of Clinical Pharmacokinetics, Medical University of Vienna, Waehringer Guertel 18-20, A-1090 Vienna, Austria. friederike.traunmueller@meduniwien.ac.at
Article em En | MEDLINE | ID: mdl-16782412
ABSTRACT
Microdialysis is an increasingly employed technique for the determination of tissue pharmacokinetics. A high-performance liquid chromatography method for the quantitative determination of caspofungin in human microdialysates with amperometric detection is described. Since microdialysis of caspofungin is performed with a 100,000 molecular mass cut-off membrane, microdialysates contain protein that was precipitated at pH 4 with acetonitrile. Addition of 1-propanol (33%, v/v) to the sample extract improved the analytical recovery to 81-89%. Caspofungin and the internal standard clarithromycin were separated isocratically on a cyanopropyl silica column using acetonitrile-0.05 M citrate (3367, v/v), adjusted to an apparent pH of 6.9, at a flow rate of 1.0 ml/min, and amperometric detection at +950 mV oxidation potential. Within-day and between-day imprecision and inaccuracy were <11%. The lower limit of quantification was 0.07 microg/ml. The method was applied to in vitro microdialysis experiments. Ringer's solution containing 1% (w/v) human albumin was used for the perfusing and surrounding medium, respectively. Albumin did not entirely prevent adsorption of caspofungin to the surface of membrane and/or tubing. When the binding-sites were saturated with albumin plus caspofungin prior to the start of sampling, the percentage of drug appearing in the microdialysate ("recovery") remained stable over the concentration range tested.
Assuntos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Peptídeos Cíclicos / Cromatografia Líquida de Alta Pressão Idioma: En Ano de publicação: 2006 Tipo de documento: Article
Buscar no Google
Base de dados: MEDLINE Assunto principal: Peptídeos Cíclicos / Cromatografia Líquida de Alta Pressão Idioma: En Ano de publicação: 2006 Tipo de documento: Article