A reliable and inexpensive EasyCD4 assay for monitoring HIV-infected individuals in resource-limited settings.
J Acquir Immune Defic Syndr ; 43(1): 23-6, 2006 Sep.
Article em En | MEDLINE | ID: mdl-16885780
ABSTRACT
Serial measurements of absolute CD4+ T-lymphocyte counts are required to initiate and gauge response to therapy and monitor disease progression. Hence, there is an urgent need to evaluate the accuracy and validity of low-cost CD4+ T-cell count assays. Tripotassium EDTA blood specimens from HIV-infected individuals were studied using a novel flow cytometric assay (EasyCD4 assay; Guava Technologies, Hayward, CA) in comparison with standard flow cytometry (FACSCount; Becton Dickinson Immunocytometry Systems, San Jose, CA). The sensitivity, specificity value by EasyCD4 assay in enumerating absolute CD4+ T-cell counts of less than 200 cells/microL were 95% and 100%, respectively. Bland-Altman analysis showed close agreement, with the EasyCD4 assay yielding CD4+ T-cell counts a mean difference of -26 cells/microL (95% confidence interval, -96 to 44 cells/microL) higher than by flow cytometry. Our data suggest that EasyCD4 assay could be a useful alternative assay to conventional flow cytometry, may be appropriate for use in resource-limited settings.
Assuntos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Infecções por HIV / Contagem de Linfócito CD4 Idioma: En Ano de publicação: 2006 Tipo de documento: Article
Buscar no Google