Your browser doesn't support javascript.
loading
Subcutaneous array with active can implantable cardioverter defibrillator configuration: a follow-up study.
Kaltman, Jonathan R; Gaynor, J William; Rhodes, Larry A; Buck, Karen; Shah, Maully J; Vetter, Victoria L; Madan, Nandini; Tanel, Ronn E.
Afiliação
  • Kaltman JR; Division of Cardiology, The Children's Hospital of Philadelphia, Philadelphia, PA 19104, USA. kaltman@gmail.com
Congenit Heart Dis ; 2(2): 125-9, 2007.
Article em En | MEDLINE | ID: mdl-18377489
ABSTRACT

BACKGROUND:

Novel nontransvenous implantable cardioverter defibrillator (ICD) configurations are sometimes required for small children and children with complex congenital heart disease at risk for sudden death. Mid- to long-term follow-up of these nontraditional implant techniques is not well known. We assessed the mid-term performance of a subcutaneous lead technique used in our practice.

METHODS:

Between July 2002 and November 2003, 4 patients (age 2.1-8.5 years, weight 13-33.3 kg, height 90-126.7 cm) received an ICD with a single-finger (n = 3) or 2-finger (n = 1) subcutaneous array with an active abdominal can and epicardial pace/sense lead. The subcutaneous tunnel was created via a subxiphoid incision using a tunneling tool within a sheath along the seventh intercostal space and extended posterior to the spine. Diagnoses included long QT syndrome (n = 2), idiopathic ventricular fibrillation (n = 1), and idiopathic dilated cardiomyopathy (n = 1). Implantable cardioverter defibrillator indications included syncope (n = 2) and cardiac arrest (n = 2).

RESULTS:

Mean follow-up was 22.3 +/- 13.9 months. During follow-up, 1 patient underwent heart transplantation and the other 3 patients underwent generator replacement secondary to a manufacturer's advisory. There was 1 appropriate and successful shock for ventricular fibrillation. This patient experienced a second episode of ventricular fibrillation that the ICD discharge failed to terminate. The arrhythmia spontaneously resolved. There were no inappropriate shocks. There was 1 instance of false detection of ventricular fibrillation because of intermittent T wave oversensing but therapy was not administered. There were no infections, lead fractures, or other complications during follow-up.

CONCLUSION:

This novel nontransvenous ICD configuration can be used safely in a select group of pediatric patients and allows for the applicability of this life-saving technology to small children at high risk for sudden cardiac death.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Desfibriladores Implantáveis Idioma: En Ano de publicação: 2007 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Desfibriladores Implantáveis Idioma: En Ano de publicação: 2007 Tipo de documento: Article