A pilot study exploring the possibility of establishing guidance levels in x-ray directed interventional procedures.
Balter, S; Miller, D L; Vano, E; Ortiz Lopez, P; Bernardi, G; Cotelo, E; Faulkner, K; Nowotny, R; Padovani, R; Ramirez, A.
Afiliação
  • Balter S; Columbia University Medical Center, 627 West 165th Street, New York, New York 10032, USA. stb2001@nyp.org
Med Phys ; 35(2): 673-80, 2008 Feb.
Article em En | MEDLINE | ID: mdl-18383689
ABSTRACT
This article summarizes the dosimetric results of an International Atomic Energy Agency coordinated research program to investigate the feasibility of adopting guidance levels for invasive coronary artery procedures. The main study collected clinical data from hospitals located in five countries. A total of 2265 coronary angiograms (CA) and 1844 percutaneous coronary interventions (PCI) were analyzed. Substudies evaluated the dosimetric performance of 14 fluoroscopes, skin dose maps obtained using film, the quality of CA procedures, and the complexity of PCI procedures. Kerma-area product (PKA) guidance levels of 50 and 125 Gy cm2 are suggested for CA and PCI procedures. These levels should be adjusted for the complexity of the procedures performed in a given institution.
Assuntos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Proteção Radiológica / Radiometria / Radiografia Intervencionista / Guias de Prática Clínica como Assunto Idioma: En Ano de publicação: 2008 Tipo de documento: Article
Buscar no Google
Base de dados: MEDLINE Assunto principal: Proteção Radiológica / Radiometria / Radiografia Intervencionista / Guias de Prática Clínica como Assunto Idioma: En Ano de publicação: 2008 Tipo de documento: Article