Your browser doesn't support javascript.
loading
[Air embolism on central venous access in a premature neonate, beware of the infusion device assembly! ]. / Embolie gazeuse sur cathéter central chez un nouveau-né prématuré, méfiez-vous de votre montage de perfusion !
Thiriez, G; Choulet, M-H; Felix, S; Afroukh, N; Nobili, F; Chamouine, A; Menget, A.
Afiliação
  • Thiriez G; Service de pédiatrie 2, CHU Saint-Jacques, 2, place Saint-Jacques, 25030 Besançon, France. gerard.thiriez@univ-fcomte.fr
Arch Pediatr ; 15(12): 1772-4, 2008 Dec.
Article em Fr | MEDLINE | ID: mdl-18976890
ABSTRACT
A fatal pulmonary air embolism, confirmed by an oriented necropsy, is described in a 25-day-old premature and small-for-gestational-age neonate. The embolism was suspected after air bubbles were detected in the infusion line. The air bubbles originated from a male Luer-lock to male Luer-lock connector, which was part of a specific assembly of the perfusion line. A solution to this problem is thereby proposed. We recommend caution in using complex infusion line assemblies because such complications are probably underestimated.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Embolia Pulmonar / Embolia Aérea / Doenças do Prematuro Idioma: Fr Ano de publicação: 2008 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Embolia Pulmonar / Embolia Aérea / Doenças do Prematuro Idioma: Fr Ano de publicação: 2008 Tipo de documento: Article