Survivorship of the Neuflex silicone implant in MCP joint replacement.
Kimani, B Mwaura; Trail, I A; Hearnden, A; Delaney, R; Nuttall, D.
Afiliação
  • Kimani BM; Wrightington Hospital, Wigan, UK.
J Hand Surg Eur Vol ; 34(1): 25-8, 2009 Feb.
Article em En | MEDLINE | ID: mdl-19270002
ABSTRACT
The Neuflex silicone metacarpophalangeal (MCP) joint replacement has previously been shown to provide improved movement, particularly flexion, when compared with the Swanson implant. In this study, we reviewed the outcome of 237 Neuflex implants in 66 patients with a follow-up of up to 7 years. Kaplan-Meier analysis using revision as the end stage revealed survival at 7 years to be 88%. If however an implant fracture, as seen radiographically, is taken as the end point, the survivorship drops to 68% at 7 years. These figures are comparable with a similar analysis for the Swanson implant. We conclude that the improved range of motion of the Neuflex implant demonstrated previously does not result in either a higher or a lower revision or implant fracture rate when compared with the Swanson implant.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Silicones / Análise de Falha de Equipamento / Prótese Articular / Articulação Metacarpofalângica Idioma: En Ano de publicação: 2009 Tipo de documento: Article