A warfarin treated patient unexpectedly presented with an elevated international normalized ratio (INR). Repeat testing in two laboratories gave conflicting results. The chromogenic assay of factor X was used to determine the correct INR result. The patient had laboratory results consistent with a dysfibrinogenemia, which prevented detection of the endpoint with a photo-optical detection system. The chromogenic assay of factor X is recommended for monitoring patients on warfarin when the INR cannot be accurately determined due to interference with the fibrin endpoint in the INR.