A flow injection system, comprising a biosensor based on a screen-printed carbon electrode containing Meldola's Blue-Reinecke salt coated with glucose dehydrogenase, for the measurement of glucose.
Anal Biochem
; 396(2): 269-74, 2010 Jan 15.
Article
em En
| MEDLINE
| ID: mdl-19766585
A biosensor for the measurement of glucose in serum has been developed, based on a screen-printed carbon electrode modified with Meldola's Blue-Reinecke salt, coated with the enzyme glucose dehydrogenase (from Bacillus sp.), and nicotinamide adenine dinucleotide coenzyme (NAD+). A cellulose acetate layer was deposited on top of the device to act as a permselective membrane. The biosensor was incorporated into a commercially available, thin-layer, amperometric flow cell operated at a potential of only +0.05 V versus Ag/AgCl. The mobile phase consisted of 0.2 M phosphate buffer (pH 7.0) containing 0.1 M potassium chloride solution, and a flow rate of 0.8 ml min(-1) was used throughout the investigation. The biosensor response was linear over the range of 0.075-30 mM glucose, with the former representing the detection limit. The precision of the system was determined by carrying out 20 repeat injections of a 5-mM glucose standard, and the calculated coefficient of variation was 3.9%. It was demonstrated that this biosensor system could be applied to the direct measurement of glucose in serum without pretreatment. Therefore, this would allow high-throughput analysis, at low cost, for this clinically important analyte.
Texto completo:
1
Base de dados:
MEDLINE
Assunto principal:
Oxazinas
/
Tiocianatos
/
Glicemia
/
Carbono
/
Técnicas Biossensoriais
/
Análise de Injeção de Fluxo
/
Glucose 1-Desidrogenase
Idioma:
En
Ano de publicação:
2010
Tipo de documento:
Article