Your browser doesn't support javascript.
loading
A comparison of retinal prosthesis electrode array substrate materials.
Weiland, James D; Humayun, Mark S; Eckhardt, Helmut; Ufer, Stefan; Laude, Lucien; Basinger, Brooke; Tai, Yu-Chong.
Afiliação
  • Weiland JD; Doheny Eye Institute, Los Angeles, California, USA.
Article em En | MEDLINE | ID: mdl-19963811
ABSTRACT
Simulations of artificial vision suggest that 1000 electrodes may be required to restore vision to individuals with diseases of the outer retina. In order to achieve such an implant, new technology is needed, since the state-of-the-art implantable neural stimulator has at most 22 contacts with neural tissue. A critical component of this system is the multi-channel, stimulating electrode array. This array must meet very challenging, competing requirements for manufacturing, integration, surgical handling, and biocompatibility. Our lab has evaluated 3 polymers as retinal prosthesis substrates polyimide, parylene, and silicone.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Próteses e Implantes / Doenças Retinianas / Eletrodos Idioma: En Ano de publicação: 2009 Tipo de documento: Article