Measurement of intraocular pressure with pressure phosphene tonometry in children.
Fan, Dorothy S P; Chiu, Thomas Y H; Congdon, Nathan; Chan, Jeffrey C W; Cheung, Eva Y Y; Lam, Dennis S C.
Afiliação
  • Fan DS; Department of Ophthalmology & Visual Sciences, The Chinese University of Hong Kong, Hong Kong Eye Hospital, Kowloon, Hong Kong, People's Republic of China.
J Pediatr Ophthalmol Strabismus ; 48(3): 167-73, 2011.
Article em En | MEDLINE | ID: mdl-20672766
ABSTRACT

PURPOSE:

To study the accuracy and acceptability of intraocular pressure (IOP) measurement by the pressure phosphene tonometer, non-contact tonometer, and Goldmann tonometer in children.

METHODS:

Fifty children (5 to 14 years old) participated in this prospective comparative study. IOP was measured with the pressure phosphene tonometer, non-contact tonometer, and Goldmann tonometer by three different examiners who were masked to the results. The children were also asked to grade the degree of discomfort from 0 to 5 (0 = no discomfort; 5 = most discomfort).

RESULTS:

The mean IOPs measured by the Goldmann tonometer, pressure phosphene tonometer, and non-contact tonometer were 15.9 mm Hg (standard deviation [SD] = 5.5 mm Hg; range 10 to 36 mm Hg), 16.0 mm Hg (SD 2.9 mm Hg; range 12 to 25 mm Hg), and 15.7 mm Hg (SD = 5.1 mm Hg; range 8 to 32 mm Hg), respectively (P = .722). The mean difference between pressure phosphene tonometer and Goldmann tonometer readings was 2.9 mm Hg and that between non-contact tonometer and Goldmann tonometer readings was 2.1 mm Hg. The 95% confidence interval of the mean difference between pressure phosphene tonometer and Goldmann tonometer readings was -1.07 and 1.19, and that between non-contact tonometer and Goldmann tonometer readings was -1.07 and 0.53. The mean discomfort ratings for the pressure phosphene tonometer, non-contact tonometer, and Goldmann tonometer were 0.6, 2.0, and 2.3, respectively (P < .001).

CONCLUSION:

Although the pressure phosphene tonometer was less accurate than the non-contact tonometer compared with Goldmann tonometer, it gave a reasonably close estimate and had a high specificity of raised IOP. In addition, measurement by the pressure phosphene tonometer is most acceptable to children. The pressure phosphene tonometer can be considered as an alternative method of IOP measurement in children.