Combined effects of implant insertion depth and alveolar bone quality on periimplant bone strain induced by a wide-diameter, short implant and a narrow-diameter, long implant.
Chou, Hsuan-Yu; Müftü, Sinan; Bozkaya, Dinçer.
Afiliação
  • Chou HY; Department of Mechanical and Industrial Engineering, Northeastern University, Boston, MA 02115, USA.
J Prosthet Dent ; 104(5): 293-300, 2010 Nov.
Article em En | MEDLINE | ID: mdl-20970535
ABSTRACT
STATEMENT OF

PROBLEM:

Strain levels in periimplant bone are affected by implant dimensions, bone quality, and implant insertion depth, resulting in different bone maintenance characteristics.

PURPOSE:

The purpose of this study was to evaluate the biomechanical response of the jaw bone to a wide-diameter, short (WDS) implant, and a narrow-diameter, long (NDL) implant for various simulated clinical scenarios. MATERIAL AND

METHODS:

The finite element method was used to evaluate periimplant bone strain distribution for 5 × 6-mm (WDS) and 3.5 × 10.7-mm (NDL) implants. A 3-dimensional segment of the mandible was constructed from a computerized tomography image of the premolar region. Occlusal force was simulated by applying a 100-N oblique load on the abutment. Bone strain distributions for 5 different implant insertion depths and 2 different levels of alveolar bone quality were evaluated.

RESULTS:

For an NDL implant, approximately 60% to 80% of the bone volume surrounding the implant was subjected to 200-1000 µstrain (µÉ›), and 15% to 35% was subjected to 1000-3000 µÉ›, regardless of the alveolar bone quality. For a WDS implant, the bone volume subjected to 1000-3000 µÉ› increased, and the bone volume subjected to 200-1000 µÉ› decreased in lower quality alveolar bone. For both implant types, bone volume experiencing strain levels less than 200 µÉ›, and/or greater than 3000 µÉ›, was predicted to be relatively small.

CONCLUSIONS:

In general, the thread design promoted relatively high strain around the thread tips, and the bone inside grooves was less strained. A more even and higher strain distribution in the periimplant bone was generated by the WDS implant as compared to the NDL implant. Regardless of the implant dimensions and simulated clinical scenarios, the development of high strain in the alveolar region was inevitable. Strain levels in periimplant bone were reduced as the insertion depth of the implant was increased.