Feasibility and radiation dose of high-pitch acquisition protocols in patients undergoing dual-source cardiac CT.
Sommer, Wieland H; Albrecht, Edda; Bamberg, Fabian; Schenzle, Jan C; Johnson, Thorsten R; Neumaier, Klement; Reiser, Maximilian F; Nikolaou, Konstatin.
Afiliação
  • Sommer WH; Department of Clinical Radiology, University of Munich, Grosshadern Campus, Marchioninstr. 15, 81377, Munich, Germany. Wieland.sommer@med.uni-muenchen.de
AJR Am J Roentgenol ; 195(6): 1306-12, 2010 Dec.
Article em En | MEDLINE | ID: mdl-21098188
ABSTRACT

OBJECTIVE:

The objective of this study was to compare image quality and radiation dose between high-pitch and established retrospectively and prospectively gated cardiac CT protocols using an Alderson-Rando phantom and a set of patients. MATERIALS AND

METHODS:

An anthropomorphic Alderson-Rando phantom equipped with thermoluminiscent detectors and a set of clinical patients underwent the following cardiac CT protocols high-pitch acquisition (pitch 3.4), prospectively triggered acquisition, and retrospectively gated acquisition (pitch 0.2). For patients with sinus rhythm below 65 beats per minute (bpm), high-pitch protocol was used, whereas for patients in sinus rhythm between 65 and 100 bpm, prospective triggering was used. Patients with irregular heart rates or heart rates of ≥ 100 bpm, were examined using retrospectively gated acquisition. Evaluability of coronary artery segments was determined, and effective radiation dose was derived from the phantom study.

RESULTS:

In the phantom study, the effective radiation dose as determined with thermoluminescent detector (TLD) measurements was lowest in the high-pitch acquisition (1.21, 3.12, and 11.81 mSv, for the high-pitch, the prospectively triggered, and the retrospectively gated acquisition, respectively). There was a significant difference with respect to the percentage of motion-free coronary artery segments (99%, 87%, and 92% for high-pitch, prospectively triggered, and retrospectively gated, respectively (p < 0.001), whereas image noise was lowest for the high-pitch protocol (p < 0.05).

CONCLUSION:

High-pitch scans have the potential to reduce radiation dose up to 61.2% and 89.8% compared with prospectively triggered and retrospectively gated scans. High-pitch protocols lead to excellent image quality when used in patients with stable heart rates below 65 bpm.