Water equivalent path length measurement in proton radiotherapy using time resolved diode dosimetry.
Med Phys
; 38(4): 2282-8, 2011 Apr.
Article
em En
| MEDLINE
| ID: mdl-21626963
PURPOSE: To verify water equivalent path length (WEPL) before treatment in proton radiotherapy using time resolved in vivo diode dosimetry. METHODS: Using a passively scattered range modulated proton beam, the output of a diode driving a fast current-to-voltage amplifier is recorded at a number of depths in a water tank. At each depth, a burst of overlapping single proton pulses is observed. The rms duration of the burst is computed and the resulting data set is fitted with a cubic polynomial. RESULTS: When the diode is subsequently set to an arbitrary depth and the polynomial is used as a calibration curve, the "unknown" depth is determined within 0.3 mm rms. CONCLUSIONS: A diode or a diode array, placed (for instance) in the rectum in conjunction with a rectal balloon, can potentially determine the WEPL at that point, just prior to treatment, with submillimeter accuracy, allowing the beam energy to be adjusted. The associated unwanted dose is about 0.2% of a typical single fraction treatment dose.