Determination of Cu2+ using poly(2-aminothiazole)/ multi-walled carbon nanotubes composite film modified glassy carbon electrodes.
Zhao, Hong; Wu, Zhijiao; Xue, Ying; Cao, Qian; He, Yujian; Li, Xiangjun; Yuan, Zhuobin.
Afiliação
  • Zhao H; College of Chemistry and Chemical Engineering, Graduate University of Chinese Academy of Sciences, Beijing 100049, China.
J Nanosci Nanotechnol ; 11(4): 3381-4, 2011 Apr.
Article em En | MEDLINE | ID: mdl-21776713
ABSTRACT
2-Aminothiazole was electropolymerized by cyclic voltammetry (CV) on the multi-walled carbon nanotubes (MWCNTs) modified glassy carbon electrode (GCE) surface. Poly(2-aminothiazole)/MWCNTs/GCE was used for determination of copper ions. The anodic peak currents of copper ions evaluated by differential pulse stripping voltammetry (DPSV) are linear with the concentrations in the range from 1.0 x 10(-7) M to 2.0 x 10(-5) M with a linear coefficiency of 0.9985. The detection limit is 2.0 x 10(-9) M calculated for a signal-to-noise ratio of 3 (S/N = 3). The proposed method was applied successfully to the determination of copper ions in drinking water, and the recovery was 96%.