Your browser doesn't support javascript.
loading
Automated determination of rifampicin in plasma samples by in-tube solid-phase microextraction coupled with liquid chromatography.
Melo, L P; Queiroz, R H C; Queiroz, M E C.
Afiliação
  • Melo LP; Departamento de Química, Faculdade de Filosofia Ciências e Letras de Ribeirão Preto, Universidade de São Paulo, Ribeirão Preto 14040-901, SP, Brazil.
J Chromatogr B Analyt Technol Biomed Life Sci ; 879(24): 2454-8, 2011 Aug 15.
Article em En | MEDLINE | ID: mdl-21778122
ABSTRACT
A sensitive and automated method is described for determination of rifampicin in plasma samples for therapeutic drug monitoring by in-tube solid-phase microextraction coupled with liquid chromatography (in-tube SPME/LC). Important factors in the optimization of in-tube SPME are discussed, such as coating type, sample pH, sample draw/eject volume, number of draw/eject cycles, and draw/eject flow rate. Analyte pre-concentrated in the polyethylene glycol phase was directly transferred to the liquid chromatographic column by percolation of the mobile phase, without carryover. The method was linear over the 0.1-100 µg/mL range, with a linear coefficient value (r(2)) of 0.998. The inter-assay precision presented coefficient of variation ≤ 1.7%. The effectiveness and practicability of the proposed method are proven by analysis of plasma samples from ageing patients undergoing therapy with rifampicin.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Rifampina / Cromatografia Líquida / Microextração em Fase Sólida Idioma: En Ano de publicação: 2011 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Rifampina / Cromatografia Líquida / Microextração em Fase Sólida Idioma: En Ano de publicação: 2011 Tipo de documento: Article