ABSTRACT
The activities of (177)Lu, (111)In and (123)I solutions have been absolutely determined using three different measurement methods. (177)Lu solution was standardized using the 4πß(PC)-γ(NaI) coincidence and 4πß(LS)-γ(NaI) live-timed anticoincidence methods. For the (111)In and (123)I solutions, besides these two mentioned methods, the coincidence sum-peak method was also applied. The measured activities results using these different methods are consistent within the evaluated experimental uncertainties demonstrating the equivalence of these methods. As an additional contribution to nuclear data, the half-lives have been determined using a well type IG12 ionization chamber.