Percutaneous interventional techniques to remove embolized silicone port catheters from heart and great vessels.
Kawata, Masahito; Ozawa, Kazuyuki; Matsuura, Takeshi; Kuroda, Masaru; Hirayama, Yasutaka; Adachi, Kazumasa; Matsuura, Akira; Sakamoto, Susumu.
Afiliação
  • Kawata M; Department of Cardiology, Akashi Medical Center, 743-33 Yagi, Ohkubo-cho, Akashi, 674-0063, Japan. kawata@amc1.jp
Cardiovasc Interv Ther ; 27(3): 196-200, 2012 Sep.
Article em En | MEDLINE | ID: mdl-22991143
Rupture of a silicon port catheter is a relatively rare complication and sometimes it is difficult to remove it. We experienced three cases of retrieval of silicone port catheters migrating into cardiac ventricle or pulmonary artery. Several devices such as a snare wire, an ablation catheter, and a basket catheter in combination with interventional guiding catheter were applied to retrieve them. These interventional techniques are applicable for retrieval of embolized vascular access port system and other catheter fragments.
Assuntos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Artéria Pulmonar / Cateterismo Venoso Central / Migração de Corpo Estranho / Embolia / Falha de Equipamento / Intervenção Coronária Percutânea / Dispositivos de Acesso Vascular / Ventrículos do Coração Idioma: En Ano de publicação: 2012 Tipo de documento: Article
Buscar no Google
Base de dados: MEDLINE Assunto principal: Artéria Pulmonar / Cateterismo Venoso Central / Migração de Corpo Estranho / Embolia / Falha de Equipamento / Intervenção Coronária Percutânea / Dispositivos de Acesso Vascular / Ventrículos do Coração Idioma: En Ano de publicação: 2012 Tipo de documento: Article