Clinical assessment of the 1/3 radius using a new desktop ultrasonic bone densitometer.
Stein, Emily M; Rosete, Fernando; Young, Polly; Kamanda-Kosseh, Mafo; McMahon, Donald J; Luo, Gangming; Kaufman, Jonathan J; Shane, Elizabeth; Siffert, Robert S.
Afiliação
  • Stein EM; Department of Medicine, Columbia University College of Physicians and Surgeons, New York, NY, USA.
Ultrasound Med Biol ; 39(3): 388-95, 2013 Mar.
Article em En | MEDLINE | ID: mdl-23312957
ABSTRACT
The objectives of this study were to evaluate the capability of a novel ultrasound device to clinically estimate bone mineral density (BMD) at the 1/3 radius. The device rests on a desktop and is portable, and permits real-time evaluation of the radial BMD. The device measures two net time delay (NTD) parameters, NTD(DW) and NTD(CW). NTD(DW) is defined as the difference between the transit time of an ultrasound pulse to travel through soft-tissue, cortex and medullary cavity, and the transit time through soft tissue only of equal overall distance. NTD(CW) is defined as the difference between the transit time of an ultrasound pulse to travel through soft-tissue and cortex only, and the transit time through soft tissue only again of equal overall distance. The square root of the product of these two parameters is a measure of the radial BMD at the 1/3 location as measured by dual-energy X-ray absorptiometry (DXA). A clinical IRB-approved study measured ultrasonically 60 adults at the 1/3 radius. BMD was also measured at the same anatomic site and time using DXA. A linear regression using NTD produced a linear correlation coefficient of 0.93 (p < 0.001). These results are consistent with previously reported simulation and in vitro studies. In conclusion, although X-ray methods are effective in bone mass assessment, osteoporosis remains one of the largest undiagnosed and under-diagnosed diseases in the world today. The research described here should enable significant expansion of diagnosis and monitoring of osteoporosis through a desktop device that ultrasonically assesses bone mass at the 1/3 radius.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Rádio (Anatomia) / Densidade Óssea Idioma: En Ano de publicação: 2013 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Rádio (Anatomia) / Densidade Óssea Idioma: En Ano de publicação: 2013 Tipo de documento: Article