Your browser doesn't support javascript.
loading
[Measurement of acetonitrile in blood and urine by head-space gas chromatography].
Luo, Ru-Xin; Zhuo, Xian-Yi; Shen, Bao-Hua.
Afiliação
  • Luo RX; School of Pharmacy, Medical College of Soochow University, Suzhou 215123, China. luoruxin1986@sina.cn
Fa Yi Xue Za Zhi ; 28(6): 422-5, 2012 Dec.
Article em Zh | MEDLINE | ID: mdl-23484321
ABSTRACT

OBJECTIVE:

To establish the method for measurement of acetonitrile in blood and urine by head-space gas chromatography.

METHODS:

DB-ALC1 (30 m x 320 microm x 1.8 microm) and DB-ALC2 (30 m x 320 microm x 1.2 microm) capillary column were used to measure the acetonitrile in blood and urine with the isopropanol as internal standard reference.

RESULTS:

The limits of detection of acetonitrile in both blood and urine were 0.5 microg/mL, with a linear range of 5-1000 microg/mL (r = 0.999).The accuracy of this method was 93.2%-98.0%. The RSD for the intra-day and inter-day were less than 3.7%.

CONCLUSION:

The method is capable for measurement analysis of acetonitrile in blood and urine.
Assuntos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Acetonitrilas / Cromatografia Gasosa Idioma: Zh Ano de publicação: 2012 Tipo de documento: Article
Buscar no Google
Base de dados: MEDLINE Assunto principal: Acetonitrilas / Cromatografia Gasosa Idioma: Zh Ano de publicação: 2012 Tipo de documento: Article