Your browser doesn't support javascript.
loading
Direct growth of vertically-oriented graphene for field-effect transistor biosensor.
Mao, Shun; Yu, Kehan; Chang, Jingbo; Steeber, Douglas A; Ocola, Leonidas E; Chen, Junhong.
Afiliação
  • Mao S; Department of Mechanical Engineering, University of Wisconsin-Milwaukee, 3200 North Cramer Street, Milwaukee, WI 53211, USA.
Sci Rep ; 3: 1696, 2013.
Article em En | MEDLINE | ID: mdl-23603871
A sensitive and selective field-effect transistor (FET) biosensor is demonstrated using vertically-oriented graphene (VG) sheets labeled with gold nanoparticle (NP)-antibody conjugates. VG sheets are directly grown on the sensor electrode using a plasma-enhanced chemical vapor deposition (PECVD) method and function as the sensing channel. The protein detection is accomplished through measuring changes in the electrical signal from the FET sensor upon the antibody-antigen binding. The novel biosensor with unique graphene morphology shows high sensitivity (down to ~2 ng/ml or 13 pM) and selectivity towards specific proteins. The PECVD growth of VG presents a one-step and reliable approach to prepare graphene-based electronic biosensors.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Transistores Eletrônicos / Imunoensaio / Técnicas Biossensoriais / Condutometria / Nanopartículas Metálicas / Ouro / Grafite Idioma: En Ano de publicação: 2013 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Transistores Eletrônicos / Imunoensaio / Técnicas Biossensoriais / Condutometria / Nanopartículas Metálicas / Ouro / Grafite Idioma: En Ano de publicação: 2013 Tipo de documento: Article