Development and validation of a stability-indicating size-indicating size-exclusion LC method for the determination of rhIFN-alpha2a in pharmaceutical formulations.
J AOAC Int
; 96(2): 324-30, 2013.
Article
em En
| MEDLINE
| ID: mdl-23767357
A size-exclusion LC method was validated for the determination of interferon-a2a (rhlFN-alpha2a) in pharmaceutical formulations without interference from human serum albumin. Chromatographic separation was performed on a BioSep-SEC-S 2000 column (300 x 7.8 mm id). The mobile phase consisted of 0.001 M monobasic potassium phosphate, 0.008 M sodium phosphate dibasic; 0.2 M sodium chloride buffer, pH 7.4, run at a gradient flow rate and using photodiode array detection at 214 nm, was used. Chromatographic separation was achieved with a retention time of 17.2 min, and the analysis was linear over the concentration range of 1.98 to 198 microg/mL (r2 = 0.9996). The accuracy was 101.39%, with bias lower than 1.67%. The LOD and LOQ were 0.87 and 1.98 microg/mL, respectively. Moreover, method validation demonstrated acceptable results for precision and robustness. The method was applied to the assessment of rhlFN-alpha2a and related proteins in biopharmaceutical dosage forms, and the content/potencies were correlated to those given by a validated RP-LC method and an in vitro bioassay. It was concluded that use of the methods in conjunction allows a great improvement in monitoring stability and QC, thereby ensuring the therapeutic efficacy of the biotechnology-derived medicine.
Buscar no Google
Base de dados:
MEDLINE
Assunto principal:
Cromatografia Líquida
/
Interferon-alfa
Idioma:
En
Ano de publicação:
2013
Tipo de documento: