Postoperative displacement of deep brain stimulation electrodes related to lead-anchoring technique.
Contarino, M Fiorella; Bot, Maarten; Speelman, Johannes D; de Bie, Rob M A; Tijssen, Marina A; Denys, Damiaan; Bour, Lo J; Schuurman, P Richard; van den Munckhof, Pepijn.
Afiliação
  • Contarino MF; Departments of †Neurology; ‡Neurosurgery; and §Psychiatry, Academic Medical Center, University of Amsterdam, Amsterdam, the Netherlands; ¶Department of Neurology, University Medical Centre Groningen, Groningen, the Netherlands.
Neurosurgery ; 73(4): 681-8; discussion 188, 2013 Oct.
Article em En | MEDLINE | ID: mdl-23842551
BACKGROUND: Displacement of deep brain stimulation (DBS) electrodes may occur after surgery, especially due to large subdural air collections, but other factors might contribute. OBJECTIVE: To investigate factors potentially contributing to postoperative electrode displacement, in particular, different lead-anchoring techniques. METHODS: We retrospectively analyzed 55 patients (106 electrodes) with Parkinson disease, dystonia, tremor, and obsessive-compulsive disorder in whom early postoperative and long-term follow-up computed tomography (CT) was performed. Electrodes were anchored with a titanium microplate or with a commercially available plastic cap system. Two independent examiners determined the stereotactic coordinates of the deepest DBS contact on early postoperative and long-term follow-up CT. The influence of age, surgery duration, subdural air volume, use of microrecordings, fixation method, follow-up time, and side operated on first was assessed. RESULTS: Subdural air collections measured on average 4.3 ± 6.2 cm. Three-dimensional (3-D) electrode displacement and displacement in the X, Y, and Z axes significantly correlated only with the anchoring method, with larger displacement for microplate-anchored electrodes. The average 3-D displacement for microplate-anchored electrodes was 2.3 ± 2.0 mm vs 1.5 ± 0.6 mm for electrodes anchored with the plastic cap (P = .030). Fifty percent of the microplate-anchored electrodes showed 2-mm or greater (potentially relevant) 3-D displacement vs only 25% of the plastic cap-anchored electrodes (P < .01). CONCLUSION: The commercially available plastic cap system is more efficient in preventing postoperative DBS electrode displacement than titanium microplates. A reliability analysis of the electrode fixation is warranted when alternative anchoring methods are used.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Complicações Pós-Operatórias / Estimulação Encefálica Profunda / Eletrodos Implantados Idioma: En Ano de publicação: 2013 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Complicações Pós-Operatórias / Estimulação Encefálica Profunda / Eletrodos Implantados Idioma: En Ano de publicação: 2013 Tipo de documento: Article