Comparison of a commercial dengue IgM capture ELISA with dengue antigen focus reduction microneutralization test and the Centers for Disease Control dengue IgM capture-ELISA.
J Virol Methods
; 195: 247-9, 2014 Jan.
Article
em En
| MEDLINE
| ID: mdl-24161816
Dengue (DENV) infection is caused by an arbovirus that is a member of the family Flaviviridae, genus Flavivirus. The diagnosis of acute dengue infection using clinical signs and symptoms can be difficult since the manifestations cannot be readily differentiated from other infections. Therefore the diagnosis of acute dengue infection depends upon laboratory assays. Dengue virus ELISAs have been designed for the detection of IgM and IgG antibodies in addition to nonstructural 1 (NS1) antigens. The InBios IgM Dengue ELISA was compared to the Antigen Focus Reduction Microneutralization Test (FRµNT90) and Centers for Disease Control Dengue IgM Capture-ELISA (CDC MAC-ELISA). The calculated sensitivity, specificity and agreement of the InBios ELISA compared to FRµNT90 and CDC MAC-ELISA was 88.7% (C.I. 81.4-93.7), 93.1% (C.I. 89.1-95.8) and 91.5% (C.I. 86.3-95.0), respectively. In summary the InBios IgM Dengue ELISA sensitivity and specificity is comparable to other commercially available IgM Capture-ELISAs.
Palavras-chave
Texto completo:
1
Base de dados:
MEDLINE
Assunto principal:
Imunoglobulina M
/
Testes de Neutralização
/
Dengue
/
Anticorpos Antivirais
Idioma:
En
Ano de publicação:
2014
Tipo de documento: