ABSTRACT
Bone ingrowth into uncemented acetabular components requires intimate cup-bone contact and rigid fixation, which can be difficult to achieve in revision hip arthroplasty. This study compares polyaxial compression locking screws with non-locked and cancellous screw constructs for acetabular cup fixation. An acetabular cup modified with screw holes to provide both compression and angular stability was implanted into a bone substitute. Coronal lever out, axial torsion and push-out tests were performed with an Instron testing machine, measuring load versus displacement. Polyaxial locking compression screws significantly improved construct stiffness compared with non-locked or cancellous screws. This increased construct stiffness will likely reduce interfacial micromotion. Further research is required to determine whether this will improve bone ingrowth in vivo and reduce cup failure.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Desenho de Prótese / Parafusos Ósseos / Artroplastia de Quadril / Prótese de Quadril / Acetábulo Idioma: En Ano de publicação: 2014 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Desenho de Prótese / Parafusos Ósseos / Artroplastia de Quadril / Prótese de Quadril / Acetábulo Idioma: En Ano de publicação: 2014 Tipo de documento: Article